DETERMINATION OF 10-20 SYSTEM ELECTRODE LOCATIONS USING MAGNETIC-RESONANCE IMAGE SCANNING WITH MARKERS

被引:82
作者
LAGERLUND, TD [1 ]
SHARBROUGH, FW [1 ]
JACK, CR [1 ]
ERICKSON, BJ [1 ]
STRELOW, DC [1 ]
CICORA, KM [1 ]
BUSACKER, NE [1 ]
机构
[1] MAYO CLIN & MAYO FDN,DEPT DIAGNOST RADIOL,ROCHESTER,MN 55905
来源
ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY | 1993年 / 86卷 / 01期
关键词
ELECTRODE COORDINATES; MRI SCANNING; SPHERICAL COORDINATE SYSTEM; TOPOGRAPHIC MAPS; 10-20; SYSTEM;
D O I
10.1016/0013-4694(93)90062-Z
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We determined locations of 33 scalp electrodes used for electroencephalographic (EEG) recording by placing markers in the positions determined by the 10-20 system and performing magnetic resonance image (MRI) scanning on volunteer subjects. Small Vaseline-filled capsules glued on the scalp with collodion produced easily delineated regions of increased signal on standard MRI head images. Measurements of each capsule's coordinates in 3 dimensions were made from MRI scans. A spherical surface was fitted through the marker positions, giving an average radius and an origin (center of sphere). The coordinate axes were rotated to ensure that electrode C(z) was on the z-axis and that the y-axis was oriented in the posterior-anterior direction. Two spherical (angular) coordinates were determined for each electrode. Spherical electrode coordinates for different subjects differed by less than 20-degrees in all cases. An average and standard deviation of the spherical coordinates were calculated for each electrode. Standard deviations of several degrees were obtained. The average spherical coordinates obtained were close to those expected on the basis of applying the 10-20 system of placement to an ideal sphere. These measurements provide data necessary for various analyses of EEG performed to help localize epileptic foci.
引用
收藏
页码:7 / 14
页数:8
相关论文
共 9 条
  • [1] GALAMBOS R, 1953, FED PROC, V12, P48
  • [2] ONLINE TRANSFORMATION OF EEG SCALP POTENTIALS INTO ORTHOGONAL SOURCE DERIVATIONS
    HJORTH, B
    [J]. ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1975, 39 (05): : 526 - 530
  • [3] LAW S K, 1991, Brain Topography, V3, P365, DOI 10.1007/BF01129639
  • [4] NUNEZ PL, 1988, FUNCTIONAL BRAIN IMAGING, P3
  • [5] CURRENT SOURCE DENSITY-ESTIMATION AND INTERPOLATION BASED ON THE SPHERICAL HARMONIC FOURIER EXPANSION
    PASCUALMARQUI, RD
    GONZALEZANDINO, SL
    VALDESSOSA, PA
    BISCAYLIRIO, R
    [J]. INTERNATIONAL JOURNAL OF NEUROSCIENCE, 1988, 43 (3-4) : 237 - 249
  • [6] SPHERICAL SPLINES FOR SCALP POTENTIAL AND CURRENT-DENSITY MAPPING
    PERRIN, F
    PERNIER, J
    BERTRAND, O
    ECHALLIER, JF
    [J]. ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1989, 72 (02): : 184 - 187
  • [7] PERRIN F, 1987, NEUROPHYSIOL, V6, P75
  • [8] AN IMPROVED METHOD FOR LOCALIZING ELECTRIC BRAIN DIPOLES
    SALU, Y
    COHEN, LG
    ROSE, D
    SATO, S
    KUFTA, C
    HALLETT, M
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1990, 37 (07) : 699 - 705
  • [9] SOURCE DERIVATION IN CLINICAL ROUTINE EEG
    WALLIN, G
    STALBERG, E
    [J]. ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1980, 50 (3-4): : 282 - 292