MICELLAR ELECTROKINETIC CAPILLARY CHROMATOGRAPHY OF PROTEINS

被引:55
作者
STREGE, MA [1 ]
LAGU, AL [1 ]
机构
[1] ELI LILLY & CO,INDIANAPOLIS,IN 46285
关键词
D O I
10.1006/abio.1993.1214
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Separations of model proteins obtained under denaturing conditions in the presence of micellar concentrations of ionic surfactants displayed high resolution and efficiency using either bare silica or C18-derivatized silica capillaries. Superior migration time reproducibility was achieved through the use of the C18-derivatized capillaries (run-to-run migration time % RSD = 0.2), relative to that obtained in bare silica capillaries (run-to-run migration time % RSD = 2.2), in the absence of buffer replenishment. The effects of surfactant concentration and pH upon the separation of a mixture of five model proteins of varying ionic and hydrophobic character were investigated, and the application of this technique to the analysis of a recombinant DNA-derived protein in fermentation broth was demonstrated. © 1994 Academic Press, Inc. All rights reserved.
引用
收藏
页码:402 / 410
页数:9
相关论文
共 58 条