REAL-TIME TRACKING OF PARAMETERS OF LUNG-MECHANICS - EMPHASIS ON ALGORITHM TUNING

被引:15
作者
AVANZOLINI, G [1 ]
BARBINI, P [1 ]
CAPPELLO, A [1 ]
CEVENINI, G [1 ]
机构
[1] UNIV SIENA, IST CHIRURG TORACICA & CARDIOVASC, SEZ BIOINGN, VIALE BRACCI, I-53100 SIENA, ITALY
来源
JOURNAL OF BIOMEDICAL ENGINEERING | 1990年 / 12卷 / 06期
关键词
algorithm tuning; Lung mechanics; modelling; online identification;
D O I
10.1016/0141-5425(90)90059-V
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We consider the problem of tracking rapid changes in the viscous and elastic properties of the respiratory system by using mouth flow and transpulmonary pressure data measured during mechanical ventilation. A recursive least-squares algorithm with adjustable compensator is used for online estimation of an R-C model of the breathing mechanics. Specific simulation experiments are presented to provide guidelines to select suitable values for the key variable, which controls the compromise between tracking ability and noise sensitivity. The results obtained confirm the critical role of the optimum tuning in relation to the noise level. Experimental results obtained from data measured on mechanically-ventilated dogs, in which respiratory distress syndrome was intravenously induced by oleic acid, demonstrate that the tuned algorithm is able to track appropriately both the viscous and elastic properties of lung mechanics. Parameter estimates are consistent with those obtained by standard and robust offine algorithms and their time course is in qualitative agreement with known physiopathological behaviour. © 1990.
引用
收藏
页码:489 / 495
页数:7
相关论文
共 28 条