INVIVO THROMBUS FORMATION ON A GUIDEWIRE DURING INTRAVASCULAR ULTRASOUND IMAGING - EVIDENCE FOR INADEQUATE HEPARINIZATION

被引:31
作者
GRAYBURN, PA [1 ]
WILLARD, JE [1 ]
BRICKNER, ME [1 ]
EICHHORN, EJ [1 ]
机构
[1] UNIV TEXAS,HLTH SCI CTR,SW MED SCH,DALLAS,TX 75235
来源
CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS | 1991年 / 23卷 / 02期
关键词
ECHOCARDIOGRAPHY; CORONARY ARTERY DISEASE; ANTICOAGULATION;
D O I
10.1002/ccd.1810230219
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
We present a case in which thrombus formation on a guidewire was visualized during intravascular ultrasound imaging despite the administration of 10,000 units of heparin sulfate prior to the procedure. Gross thrombus formation was confirmed upon removal of the guidewire. The activated clotting time was found to be 191 s, suggesting heparin resistance. This report graphically illustrates the potential thrombogenicity associated with intravascular guidewire manipulation during procedures such as intravascular ultrasound imaging and percutaneous transluminal coronary angioplasty. Moreover, this case suggests that adequate heparinization should be routinely verified by measuring the activated clotting time prior to introducing a guidewire into the arterial system.
引用
收藏
页码:141 / 143
页数:3
相关论文
共 8 条
[1]  
BULL BS, 1975, J THORAC CARDIOV SUR, V69, P674
[2]   ANGIOGRAPHIC AND CLINICAL PREDICTORS OF ACUTE CLOSURE AFTER NATIVE VESSEL CORONARY ANGIOPLASTY [J].
ELLIS, SG ;
ROUBIN, GS ;
KING, SB ;
DOUGLAS, JS ;
WEINTRAUB, WS ;
THOMAS, RG ;
COX, WR .
CIRCULATION, 1988, 77 (02) :372-379
[3]   ACUTE CORONARY-OCCLUSION OCCURRING AFTER SUCCESSFUL PERCUTANEOUS TRANS-LUMINAL CORONARY ANGIOPLASTY - TEMPORAL RELATIONSHIP TO DISCONTINUATION OF ANTICOAGULATION [J].
GABLIANI, G ;
DELIGONUL, U ;
KERN, MJ ;
VANDORMAEL, M .
AMERICAN HEART JOURNAL, 1988, 116 (03) :696-700
[4]   ARTERIAL-WALL CHARACTERISTICS DETERMINED BY INTRAVASCULAR ULTRASOUND IMAGING - AN INVITRO STUDY [J].
GUSSENHOVEN, EJ ;
ESSED, CE ;
LANCEE, CT ;
MASTIK, F ;
FRIETMAN, P ;
VANEGMOND, FC ;
REIBER, J ;
BOSCH, H ;
VANURK, H ;
ROELANDT, J ;
BOM, N .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1989, 14 (04) :947-952
[5]   IMPORTANCE OF ADEQUATE HEPARIN DOSAGE IN ARTERIAL ANGIOPLASTY IN A PORCINE MODEL [J].
HERAS, M ;
CHESEBRO, JH ;
PENNY, WJ ;
BAILEY, KR ;
LAM, JYT ;
HOLMES, DR ;
REEDER, GS ;
BADIMON, L ;
FUSTER, V .
CIRCULATION, 1988, 78 (03) :654-660
[6]  
MILLER RD, 1986, ANESTHESIOLOGY, V64, P504
[7]   ADEQUATE HEPARINIZATION DURING PTCA - ASSESSMENT USING ACTIVATED CLOTTING TIMES [J].
OGILBY, JD ;
KOPELMAN, HA ;
KLEIN, LW ;
AGARWAL, JB .
CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1989, 18 (04) :206-209
[8]   CORONARY-ARTERY IMAGING WITH INTRAVASCULAR HIGH-FREQUENCY ULTRASOUND [J].
POTKIN, BN ;
BARTORELLI, AL ;
GESSERT, JM ;
NEVILLE, RF ;
ALMAGOR, Y ;
ROBERTS, WC ;
LEON, MB .
CIRCULATION, 1990, 81 (05) :1575-1585