VERTICAL X-RAY CONFINEMENT APPLICABLE TO A TIME RESOLVED XAFS METHOD

被引:4
作者
LIVINS, P [1 ]
THIEL, DJ [1 ]
STERN, EA [1 ]
LEWIS, A [1 ]
机构
[1] CORNELL UNIV,DEPT APPL PHYS,ITHACA,NY 14853
关键词
D O I
10.1016/0168-9002(90)90070-M
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We present results of focusing X-rays in a configuration suitable for a time-resolved X-ray absorption fine structure (XAFS) measurement in the microsecond range, an improvement by two orders of magnitude over previous techniques at synchrotron sources. The approach employs a new method of focusing X-rays by confinement and combines it with flow techniques from laser technology. The inexpensive focusing method provided a line focus of about 10 μm with intensity gains of at least 9, which produced quality XAFS spectra of a 58 millimolar solution of zinc sulfate in a fast flowing jet stream. © 1990.
引用
收藏
页码:250 / 252
页数:3
相关论文
共 8 条
[1]   MICROCOMPUTER-CONTROLLED SYSTEM USED IN TIME-RESOLVED EXAFS TECHNIQUE [J].
LIU, WH ;
WANG, XF ;
TENG, TY ;
HUANG, HW .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1983, 54 (12) :1653-1656
[2]  
LIU WH, 1983, REV SCI INSTRUM, V54, P1488
[3]   ELECTRONICS AND SAMPLE CELL FOR THE COLLECTION OF TIME-RESOLVED X-RAY ABSORPTION-SPECTRA [J].
MILLS, DM ;
POLLOCK, V ;
LEWIS, A ;
HAROOTUNIAN, A ;
HUANG, J .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1984, 222 (1-2) :351-354
[5]  
SAIGO S, 1986, J PHYSIQUE C, V8, P555
[6]   SIMPLE METHOD FOR FOCUSING X-RAYS USING TAPERED CAPILLARIES [J].
STERN, EA ;
KALMAN, Z ;
LEWIS, A ;
LIEBERMAN, K .
APPLIED OPTICS, 1988, 27 (24) :5135-5139
[7]   FOCUSING OF SYNCHROTRON RADIATION USING TAPERED GLASS-CAPILLARIES [J].
THIEL, DJ ;
STERN, EA ;
BILDERBACK, DH ;
LEWIS, A .
PHYSICA B, 1989, 158 (1-3) :314-316
[8]   TIME RESOLVED X-RAY ABSORPTION-SPECTROSCOPY [J].
TOURILLON, G ;
DARTYGE, E ;
FONTAINE, A ;
TOLENTINO, H .
PHYSICA B, 1989, 158 (1-3) :287-290