A PRACTICAL METHOD FOR POSITION-DEPENDENT COMPTON-SCATTER CORRECTION IN SINGLE PHOTON-EMISSION CT

被引:473
作者
OGAWA, K
HARATA, Y
ICHIHARA, T
KUBO, A
HASHIMOTO, S
机构
[1] SHOWA UNIV,SCH DENT,DEPT DENT RADIOL,TOKYO 145,JAPAN
[2] TOSHIBA NASU WORKS,TOCHIGI 32926,JAPAN
[3] KEIO UNIV,SCH MED,DEPT RADIOL,TOKYO 160,JAPAN
关键词
ENERGY-WEIGHTED ACQUISITION; UNIFIED RECONSTRUCTION ALGORITHM; INVERSE MONTE-CARLO; COMPUTED-TOMOGRAPHY; SPATIAL-DISTRIBUTION; ATTENUATION CORRECTION; SPECT; QUANTIFICATION; COMPENSATION; SUBTRACTION;
D O I
10.1109/42.97591
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
A new practical method was proposed to subtract the count of scattered photons from that acquired with a photopeak window at each pixel in each planar image of SPECT. The subtraction is carried out using two sets of data: one set is acquired with a main window centered at photopeak energy and the other is acquired with two subwindows on both sides of the main window. The scattered photons included in the main window are estimated from the counts acquired with the sub-windows and then they are subtracted from the count acquired with the main window. Since the subtraction is performed at each pixel in each planar image, the proposed method has the potential to be more precise than conventional methods. For three different activity distributions in cylinder phantoms, simulation tests gave good agreement between the activity distributions reconstructed from unscattered photons and those from the corrected data.
引用
收藏
页码:408 / 412
页数:5
相关论文
共 30 条
[1]  
AXELSSON B, 1984, J NUCL MED, V25, P490
[2]  
BECK JW, 1982, IEEE T NUCL SCI, V29, P506, DOI 10.1109/TNS.1982.4335896
[4]   ATTENUATION CORRECTION AND INCOMPLETE PROJECTION IN SINGLE PHOTON-EMISSION COMPUTED-TOMOGRAPHY [J].
CHANG, LT .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1979, 26 (02) :2780-2789
[5]  
DEVITO RP, 1989, J NUCL MED, V30, P2029
[6]   INTEGRAL-TRANSPORT METHOD FOR COMPTON-SCATTER CORRECTION IN EMISSION COMPUTED-TOMOGRAPHY [J].
EGBERT, SD ;
MAY, RS .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1980, 27 (01) :543-548
[7]   SCATTER DETECTION IN SPECT IMAGING - DEPENDENCE ON SOURCE DEPTH, ENERGY, AND ENERGY WINDOW [J].
FLOYD, CE ;
JASZCZAK, RJ ;
COLEMAN, RE .
PHYSICS IN MEDICINE AND BIOLOGY, 1988, 33 (09) :1075-1081
[8]  
FLOYD CE, 1985, J NUCL MED, V26, P403
[9]   MONTE-CARLO EVALUATION OF COMPTON SCATTER SUBTRACTION IN SINGLE PHOTON-EMISSION COMPUTED-TOMOGRAPHY [J].
FLOYD, CE ;
JASZCZAK, RJ ;
HARRIS, CC ;
GREER, KL ;
COLEMAN, RE .
MEDICAL PHYSICS, 1985, 12 (06) :776-778
[10]   INVERSE MONTE-CARLO - A UNIFIED RECONSTRUCTION ALGORITHM FOR SPECT [J].
FLOYD, CE ;
JASZCZAK, RJ ;
COLEMAN, RE .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1985, 32 (01) :779-785