ULTRASONIC 3-DIMENSIONAL RECONSTRUCTION - IN-VITRO AND IN-VIVO VOLUME AND AREA MEASUREMENT

被引:39
作者
HODGES, TC [1 ]
DETMER, PR [1 ]
BURNS, DH [1 ]
BEACH, KW [1 ]
STRANDNESS, DE [1 ]
机构
[1] UNIV WASHINGTON,MED CTR,INST AFFILIAT,DEPT SURG,VASC SURG SECT,SEATTLE,WA 98195
关键词
COMPUTER-ASSISTED IMAGE PROCESSING; ULTRASONOGRAPHY; HUMAN; IN VITRO; VASCULAR PATENCY;
D O I
10.1016/0301-5629(94)90029-9
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
This study validates the use of an ultrasound three-dimensional reconstruction system to measure phantom and blood conduit geometry. Independently determined uniform and stenotic phantom dimensions are compared with reconstruction-based measurements. Lower extremity saphenous vein bypass graft reconstructions were performed to demonstrate clinical application. Uniform phantom independent and reconstructed volume correlation was high (r = 0.989), the average volume difference was 4.68 mm(3) and the average area difference was 0.4 mm(2). An in vitro 28% diameter reduction was detected. Stenotic bypass graft segment volume was 795 mm(3); following successful angioplasty the volume increased to 1419 mm(3). Advantages of this technique are its accuracy, the luminal information it provides and the absence of mechanical arm or acoustic transmitter limitations. We are exploring the possibility that measurement of luminal change over time may allow stenosis detection prior to hemodynamic disturbance, in an ongoing clinical saphenous vein bypass graft surveillance study.
引用
收藏
页码:719 / 729
页数:11
相关论文
共 15 条
[1]   MEASUREMENT IN MEDICINE - THE ANALYSIS OF METHOD COMPARISON STUDIES [J].
ALTMAN, DG ;
BLAND, JM .
JOURNAL OF THE ROYAL STATISTICAL SOCIETY SERIES D-THE STATISTICIAN, 1983, 32 (03) :307-317
[2]   DURABILITY OF VEIN GRAFT REVISION - THE OUTCOME OF SECONDARY PROCEDURES [J].
BANDYK, DF ;
BERGAMINI, TM ;
TOWNE, JB ;
SCHMITT, DD ;
SEABROOK, GR .
JOURNAL OF VASCULAR SURGERY, 1991, 13 (02) :200-210
[3]   WORK IN PROGRESS - COMMON CAROTID-ARTERY CONTOURS RECONSTRUCTED IN 3 DIMENSIONS FROM PARALLEL ULTRASONIC IMAGES [J].
BLANKENHORN, DH ;
CHIN, HP ;
STRIKWERDA, S ;
BAMBERGER, J ;
HESTENES, JD .
RADIOLOGY, 1983, 148 (02) :533-537
[4]   B-MODE SONOGRAPHY OF BLOOD-CLOTS [J].
COELHO, JCU ;
SIGEL, B ;
RYVA, JC ;
MACHI, J ;
RENIGERS, SA .
JOURNAL OF CLINICAL ULTRASOUND, 1982, 10 (07) :323-327
[5]  
DETMER PR, 1994, IN PRESS ULTRASOUND
[6]  
King D L, 1992, J Am Soc Echocardiogr, V5, P569
[7]   3-DIMENSIONAL SPATIAL REGISTRATION AND INTERACTIVE DISPLAY OF POSITION AND ORIENTATION OF REAL-TIME ULTRASOUND IMAGES [J].
KING, DL ;
KING, DL ;
SHAO, MYC .
JOURNAL OF ULTRASOUND IN MEDICINE, 1990, 9 (09) :525-532
[8]  
KING DL, 1991, J ULTRAS MED, V10, P77
[9]  
Kitney R I, 1989, Int J Card Imaging, V4, P135, DOI 10.1007/BF01745143
[10]   DUPLEX SCANNING FOR DIAGNOSIS OF AORTOILIAC AND FEMOROPOPLITEAL DISEASE - A PROSPECTIVE-STUDY [J].
KOHLER, TR ;
NANCE, DR ;
CRAMER, MM ;
VANDENBURGHE, N ;
STRANDNESS, DE .
CIRCULATION, 1987, 76 (05) :1074-1080