ABSORBANCE LINEARITY AND REPEATABILITY IN CYLINDRICAL INTERNAL REFLECTANCE FT-IR SPECTROSCOPY OF LIQUIDS

被引:26
作者
DOYLE, WM
机构
[1] Axiom Analytical Inc, , CA
关键词
D O I
10.1366/0003702904086001
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
This paper reports the results of a program aimed at eliminating the sources of performance variability and excessive nonlinearity that have limited the effectiveness of the cylindrical internal reflectance technique for use in quantitative FT-IR analysis. After discussing the various optical factors that can affect performance, it describes a new adjustment-free optical design which promises to eliminate those factors which lead to data unpredictability. Finally, it summarizes the results of a series of experiments carried out to test the effectiveness of the new design. These confirm the achievement of three significant improvements: an enhancement of band strengths by factors of typically 2.5 to 3, significantly improved absorbance linearity, and a high degree of data repeatability under varied experimental conditions.
引用
收藏
页码:50 / 58
页数:9
相关论文
共 7 条
[1]   CONSISTENCY IN CIRCLE CELL FT-IR ANALYSIS OF AQUEOUS-SOLUTIONS [J].
BRAUE, EH ;
PANNELLA, MG .
APPLIED SPECTROSCOPY, 1987, 41 (06) :1057-1067
[2]  
HANSEN DL, 1987, SPROUSE COLLECTION I
[3]  
Harrick N. J., 1967, INTERNAL REFLECTION
[4]  
HIRSCHFELD T, 1979, FOURIER TRANSFORM IN, V2, P215
[5]  
RAMBAUSKE WR, 1974, Patent No. 3827059
[6]  
WILKS P, 1989, IND RES DEV SEP, P132
[7]  
10 HARR SCI CORP DAT