IN-VIVO COMPARISON BETWEEN 2-TIP PRESSURE TRANSDUCER SYSTEMS

被引:13
作者
AUBERT, AE
VROLIX, M
DEGEEST, H
VANDEWERF, F
机构
[1] Department of Cardiology, University Hospital Gasthuisberg, Catholic University Leuven
来源
INTERNATIONAL JOURNAL OF CLINICAL MONITORING AND COMPUTING | 1995年 / 12卷 / 02期
关键词
TIP CATHETER; PRESSURE TRANSDUCER; LEFT VENTRICULAR PRESSURE; CROSS CORRELATION;
D O I
10.1007/BF01142487
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Experimental findings are presented of an in vivo comparison between a Sentron catheter and another tip transducer manometer: a Millar microtip catheter. Both catheters have been used simultaneously in the left ventricle of dogs. Pressure variations were elicited by drug infusion. Pressure values and derivatives obtained from both systems were compared. A cross correlation between episodes of the two pressures was computed. Results from this study showed good correlation between left ventricular systolic pressure measured with both manometers (R = 0.992, p < 0.0001), end-diastolic pressure (R = 0.809, p < 0.0001) and between first derivatives: positive derivative (R = 0.993, p < 0.0001) and negative (R = 0.634, p < 0.0001). The mean cross correlation between both pressure signals was 0.61 +/- 0.04. In the frequency domain no statistical difference was found between the location of the maxima of the peaks. It is concluded that a Sentron manometer can be a valid alternative, at a reasonable price, to a cheaper, though less accurate fluid filled catheter and a more expensive 'golden standard' microtip catheter.
引用
收藏
页码:77 / 83
页数:7
相关论文
共 10 条
[1]  
Marey E.J., Physiologie médicale de la circulation du sang, (1863)
[2]  
Grunbaum O.F., On a new method of recording alterations of pressure, J Physiol, 22, pp. 49-51, (1897)
[3]  
Laurens P., Bouchard F., Cornu C., Baculard P., Soulie P., Bruits et pressions cardiovasculaires enregistrés in situ à l'aide d'un micromanomètre, Arch Mal Coeur, 52, pp. 121-9, (1959)
[4]  
Millar H.D., Baker L.E., A stable ultraminiature catheter-tip pressure transducer, Med Biol Eng, 11, pp. 86-9, (1973)
[5]  
Gould K., Treuholme S., Kennedy J., In vivo comparison of catheter manometer systems with the catheter-tip micromanometer, J Appl Physiol, 34, pp. 263-7, (1973)
[6]  
Bendat J.S., Piersol A.G., Random data: analysis and measurements procedures, pp. 28-31, (1971)
[7]  
Geddes L.A., Fundamentals of blood pressure transducers, Indwelling and implantable pressure transducers, pp. 5-9, (1977)
[8]  
Courtois M., Kovacs S.J., Ludbrook P.A., The transmitral pressure-flow velocity relationship: The importance of regional pressure gradients in the left ventricle, Circulation, 78, pp. 661-71, (1988)
[9]  
Hellige G., Recording of ventricular pressure by conventional catheter manometer systems. I. Minimal requirements of blood pressure recording systems and estimation of frequency response characteristics, Basic Res Cardiol, 71, pp. 319-36, (1975)
[10]  
Aubert A.E., Denys B.G., Meno F., Reddy P.S., Investigation of gallop sounds in dogs by quantitative phonocardiography and digital frequency analysis, Circulation, 71, pp. 987-93, (1985)