The performances of two alternative sample introduction methods for use with Inductively Coupled Plasma-Optical Emission Spectroscopy (ICP-OES) and Inductively Coupled Plasma-Mass Spectrometry (ICP-MS) have been evaluated for the analysis of the same sample material. The laser ablation sample introduction system is based on a Nd:YAG laser to which an x-y-z translational sampling head had been added. A study has been made of a number of parameters which affect the performance of the system to find the optimum operating conditions. The slurry introduction system involved aspirating the slurries into the plasma using a "de Galan" nebuliser and a "Scott-type" spray chamber arrangement. A study has been made of the parameters which control the production of stable homogeneous slurries. Initial particle size measurements have been carried out on the slurried samples to show how this affects this method of sample introduction. Results are presented for the analysis of a South African reference material rock sample (SARM 5) by ICP-OES with both laser ablation and slurry nebulisation sample introduction and some preliminary results for the analysis by ICP-MS with laser ablation introduction. Semi-quantitative results are obtained for laser ablation ICP-OES as only one matrix matched standard is used. However, the agreement between the results obtained for slurry nebulisation and the certificate value is poor. It is suggested by comparison with previous studies that this may be due to particle size effects. Encouraging results were obtained for the determination of trace elements by laser ablation ICP-MS. © 1990 Springer-Verlag.
引用
收藏
页码:284 / 289
页数:6
相关论文
共 25 条
[1]  
Abercrombie F. N., 1978, APPLICATIONS INDUCTI, P121
[2]   LASER ABLATION OF SOLIDS FOR ELEMENTAL ANALYSIS BY INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
ARROWSMITH, P .
ANALYTICAL CHEMISTRY, 1987, 59 (10) :1437-1444
[3]  
BABINGTON R.S., 1969, Method of atomizing liquid in a mono-dispersed spray, Patent No. [3,421,692, 3421692]
[4]   LASER VAPORIZATION OF SOLID METAL SAMPLES INTO AN INDUCTIVELY COUPLED PLASMA [J].
CARR, JW ;
HORLICK, G .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1982, 37 (01) :1-15
[5]  
DARKE SA, 1987, AERE R12821 UK AT EN
[6]   DIRECT ANALYSIS OF SOIL SUSPENSIONS BY INDUCTIVELY COUPLED PLASMA-ATOMIC EMISSION-SPECTROMETRY FOR DETERMINATION OF TOTAL METALS [J].
DICK, WA ;
PAGE, JR ;
JEWELL, KE .
SOIL SCIENCE, 1985, 139 (03) :211-218
[7]  
DITTRICH K, 1987, FRESEN Z ANAL CHEM, V328, P330
[8]   DIRECT ATOMIC SPECTROMETRIC ANALYSIS BY SLURRY ATOMIZATION .7. ANALYSIS OF COAL USING INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
EBDON, L ;
FOULKES, ME ;
PARRY, HGM ;
TYE, CT .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1988, 3 (06) :753-761
[9]   PARTICLE-SIZE EFFECTS ON KAOLIN SLURRY ANALYSIS BY INDUCTIVELY COUPLED PLASMA-ATOMIC EMISSION-SPECTROMETRY [J].
EBDON, L ;
COLLIER, AR .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1988, 43 (4-5) :355-369
[10]   A STUDY OF PNEUMATIC NEBULIZATION SYSTEMS FOR INDUCTIVELY COUPLED PLASMA EMISSION-SPECTROMETRY [J].
EBDON, L ;
CAVE, MR .
ANALYST, 1982, 107 (1271) :172-178