DETECTION OF HIGH-MASS BIOMOLECULES IN FOURIER-TRANSFORM ION-CYCLOTRON RESONANCE MASS-SPECTROMETRY - THEORETICAL AND EXPERIMENTAL INVESTIGATIONS

被引:39
作者
SOLOUKI, T [1 ]
GILLIG, KJ [1 ]
RUSSELL, DH [1 ]
机构
[1] TEXAS A&M UNIV,DEPT CHEM,COLLEGE STN,TX 77843
关键词
D O I
10.1021/ac00081a037
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A simple modification to an ion cyclotron resonance (ICR) cell for detection of high-mass ions with large spatial and kinetic energy distributions is described. The modification consists of a copper wire positioned inside the ICR cell running parallel to the magnetic field lines. The wire sets as an ion guide to position the ions at the exact center of the ion cell. Ion trajectories are calculated using SIMION, and the predictions based on the calculations are compared with the experimental results. The results presented in this paper show that the ion-guide ICR cell can be used to detect singly charged biomolecule ions of up to 157 000 Da.
引用
收藏
页码:1583 / 1587
页数:5
相关论文
共 33 条
[1]   ELECTRON COLLISION STUDIES WITH TRAPPED POSITIVE IONS [J].
BAKER, FA ;
HASTED, JB .
PHILOSOPHICAL TRANSACTIONS OF THE ROYAL SOCIETY OF LONDON SERIES A-MATHEMATICAL AND PHYSICAL SCIENCES, 1966, 261 (1115) :33-+
[2]  
BEAVIS RC, 1991, CHEM PHYS LETT, V181, P402
[3]   ULTRAHIGH-RESOLUTION MATRIX-ASSISTED LASER-DESORPTION IONIZATION OF SMALL PROTEINS BY FOURIER-TRANSFORM MASS-SPECTROMETRY [J].
CASTORO, JA ;
WILKINS, CL .
ANALYTICAL CHEMISTRY, 1993, 65 (19) :2621-2627
[4]   MATRIX-ASSISTED LASER DESORPTION IONIZATION USING A LIQUID MATRIX - FORMATION OF HIGH-MASS CLUSTER IONS FROM PROTEINS [J].
CHAN, TWD ;
COLBURN, AW ;
DERRICK, PJ .
ORGANIC MASS SPECTROMETRY, 1992, 27 (01) :53-56
[5]   COLLISION CROSS-SECTIONS FOR PROTEIN IONS [J].
COVEY, T ;
DOUGLAS, DJ .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 1993, 4 (08) :616-623
[6]  
DAHL DA, 1988, SIMON PC PS2 V 4 0
[7]  
ENS W, 1991, RAPID COMMUN MASS SP, V5, P177
[8]   ENHANCED MASS RESOLVING POWER, SENSITIVITY, AND SELECTIVITY IN LASER-DESORPTION FOURIER-TRANSFORM ION-CYCLOTRON RESONANCE MASS-SPECTROMETRY BY ION AXIALIZATION AND COOLING [J].
GUAN, SH ;
WAHL, MC ;
WOOD, TD ;
MARSHALL, AG .
ANALYTICAL CHEMISTRY, 1993, 65 (13) :1753-1757
[9]   AXIAL AND RADIAL ION CLOUD COMPRESSION - COUPLING OF MAGNETRON AND CYCLOTRON MOTION TO AXIAL MOTION IN A SEGMENTED CUBIC FOURIER-TRANSFORM ION-CYCLOTRON RESONANCE ION TRAP [J].
GUAN, SH ;
XIANG, XZ ;
MARSHALL, AG .
INTERNATIONAL JOURNAL OF MASS SPECTROMETRY AND ION PROCESSES, 1993, 124 (01) :53-67
[10]   FIELD-CORRECTED ION CELL FOR ION-CYCLOTRON RESONANCE [J].
HANSON, CD ;
CASTRO, ME ;
KERLEY, EL ;
RUSSELL, DH .
ANALYTICAL CHEMISTRY, 1990, 62 (05) :520-526