CLINICAL EVALUATION OF A DISPOSABLE ARTIFICIAL KIDNEY

被引:9
作者
KULATILAKE, AE
VICKERS, J
SHACKMAN, R
机构
[1] Department of Surgery, Royal Postgraduate Medical School, Hammersmith Hospital
来源
BMJ-BRITISH MEDICAL JOURNAL | 1969年 / 3卷 / 5668期
关键词
D O I
10.1136/bmj.3.5668.447
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The in-vivo performance of a new disposable parallel flow countercurrent type of artificial kidney has been compared with that of the modified four-layered Kiil dialyser. The dialysance of urea and endogenous creatinine in the former was significantly better than in the latter for similar blood flow rates, and, moreover, unlike the dialysance with the Kiil dialyser it continued to improve over 12 hours. Among 100 disposable artificial kidneys tested no failure occurred from blood leaks despite the use of high negative pressures in the dialysate compartment. No pyrogen reactions were observed. These new artificial kidneys were ready for use in less than one-seventh of the time taken to prepare the conventional Kiil dialyser. Other advantages are their low priming volumes, the lack of a pump in the blood line, and the reduced risk of infection. © 1969, British Medical Journal Publishing Group. All rights reserved.
引用
收藏
页码:447 / +
页数:1
相关论文
共 13 条
[1]  
ALWALL N, 1968, DIALYSIS RENAL TRANS, P18
[2]  
BERLYNE GM, 1965, EUROPEAN DIALYSIS TR, P285
[3]  
COLE JJ, 1962, T AM SOC ART INT ORG, V8, P209
[4]  
COLE JJ, 1963, T AM SOC ART INT ORG, V9, P67
[5]  
FREEMAN RB, 1964, T AM SOC ART INT ORG, V10, P174
[6]  
KIIL F, 1960, Acta Chir Scand Suppl, VSuppl 253, P142
[7]  
KOLFF WJ, 1956, J LAB CLIN MED, V47, P969
[8]  
Kolff WJ, 1944, ACTA MED SCAND, V117, P121
[9]  
KOLFF WJ, 1943, GENESK GIDS, V5, P21
[10]  
PAPADIMITRIOUS M, 1969, BIO MEDICAL ENGINEER