FORWARD-DIRECTED BREMSSTRAHLUNG OF 10-MEV TO 30-MEV ELECTRONS INCIDENT ON THICK TARGETS OF AL AND PB

被引:56
作者
FADDEGON, BA [1 ]
ROSS, CK [1 ]
ROGERS, DWO [1 ]
机构
[1] NATL RES COUNCIL CANADA, INST NATL MEASUREMENT STANDARDS, OTTAWA K1A 0R6, ONTARIO, CANADA
关键词
absolute thick-target bremsstrahlung bremsstrahlung Nal spectrometry; radiotherapy Monte Carlo;
D O I
10.1118/1.596560
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Bremsstrahlung spectra from thick targets of Al and Pb have been measured absolutely (photons per incident electron) along the beam axis for electrons of 10-, 15-, 20-, 25-, and 30-MeV incident energy. The spectra have a 220-keV low-energy cutoff. The targets were cylinders with nominal thicknesses of 110% of the electron CSDA range. A thin transmission detector, calibrated against a toroidal current monitor, was placed upstream of the target to measure the beam current. The spectrometer was a 20-cm diameter by 25-cm-long cylindrical Nal detector. Measured spectra were corrected for pile-up, background, detector response, detector efficiency, attenuation in materials between the target and detector and the collimator effect. Spectra were calculated using the EGS4 Monte Carlo system for simulating the radiation transport. The simulation model included the small amount of material upstream of the target. This material contributed about 40% of the spectrum, but its presence or absence had little effect on the calculated bremsstrahlung yield. The shapes of the measured and calculated spectra were in excellent agreement. The ratio of the total number of photons in each measured spectrum to those in the corresponding calculated spectrum varied from 0.97 + 0.06 to 1.12 + 0.06, depending largely on the atomic number of the target. Absolute spectral measurements in the literature agreed with our calculations of spectral shape but showed a range of ± 30% in the number of photons per incident electron relative to the calculated values, which is contrary to our result. © 1990, American Association of Physicists in Medicine. All rights reserved.
引用
收藏
页码:773 / 785
页数:13
相关论文
共 26 条
[1]   COLLIMATOR-PRODUCED ENERGY DEGRADATION OF MEGAVOLTAGE X-RAYS [J].
AMOLS, HI ;
RUBIN, A ;
PECK, RA .
MEDICAL PHYSICS, 1984, 11 (04) :521-524
[2]  
[Anonymous], [No title captured]
[3]   APPLICATION OF A LAPLACE TRANSFORM PAIR MODEL FOR HIGH-ENERGY X-RAY SPECTRAL RECONSTRUCTION [J].
ARCHER, BR ;
ALMOND, PR ;
WAGNER, LK .
MEDICAL PHYSICS, 1985, 12 (05) :630-633
[4]  
BIELAJEW AF, 1987, NUCL INSTRUM METH B, V18, P165
[5]  
BIELAJEW AF, 1989, PIRS0203 NAT RES COU
[6]  
FADDEGON BA, IN PRESS NUCL INS B
[7]  
FADDEGON BA, UNPUB NUCL INSTRUM A
[8]  
Jupiter C., 1967, NASASP169 NASASP169, P249
[9]   BREMSSTRAHLUNG CROSS-SECTION FORMULAS AND RELATED DATA [J].
KOCH, HW ;
MOTZ, JW .
REVIEWS OF MODERN PHYSICS, 1959, 31 (04) :920-955
[10]   MEASUREMENT OF BREMSSTRAHLUNG SPECTRA FROM 25 MEV ELECTRONS ON TA AS A FUNCTION OF RADIATOR THICKNESS AND EMISSION ANGLE [J].
LAMBERT, RP ;
JURY, JW ;
SHERMAN, NK .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH, 1983, 214 (2-3) :349-360