A HIGH COUNTING RATE PARALLEL-PLATE GAS PROPORTIONAL COUNTER FOR FLUORESCENCE EXTENDED X-RAY-ABSORPTION FINE-STRUCTURE STUDIES

被引:6
作者
SMITH, AD
PADMORE, HA
BUKSH, PA
机构
[1] Daresbury Laboratory, Warrington
关键词
D O I
10.1063/1.1142622
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Gas proportional counters (GPCs) are well suited for fluorescence extended x-ray-absorption fine structure (EXAFS) experiments in the soft x-ray (< 4 keV) region. They have a near unity detection efficiency and offer good energy resolution. Of the different types of GPC, the parallel-plate device is an attractive option for use with synchrotron radiation beam lines where its intrinsically high count rate permits optimum use of the high photon flux available. The development of a parallel-plate gas proportional counter for use on soft x-ray beam lines at the Daresbury synchrotron source is described. This device is demonstrated to have a counting rate in excess of 1 MHz, and data are presented on its performance including its energy resolution, long-term gain stability, and sample EXAFS spectra.
引用
收藏
页码:837 / 841
页数:5
相关论文
共 11 条
[1]   THE SUPPRESSION OF DESTRUCTIVE SPARKS IN PARALLEL PLATE PROPORTIONAL-COUNTERS [J].
COCKSHOTT, RA ;
MASON, IM .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1984, 31 (01) :107-110
[2]   FLUORESCENCE DETECTION OF EXAFS - SENSITIVITY ENHANCEMENT FOR DILUTE SPECIES AND THIN-FILMS [J].
JAKLEVIC, J ;
KIRBY, JA ;
KLEIN, MP ;
ROBERTSON, AS ;
BROWN, GS ;
EISENBERGER, P .
SOLID STATE COMMUNICATIONS, 1977, 23 (09) :679-682
[3]  
JONES AO, 1992, REV SCI INS, V63
[4]   MONOCHROMATOR AND BEAMLINE FOR SOFT-X-RAY STUDIES IN THE PHOTON ENERGY-RANGE 500 EV-5 KEV [J].
MACDOWELL, AA ;
WEST, JB ;
GREAVES, GN ;
VANDERLAAN, G .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1988, 59 (06) :843-852
[5]   THE EXOSAT IMAGING X-RAY-DETECTORS [J].
MASON, IM ;
BRANDUARDIRAYMONT, G ;
CULHANE, JL ;
CORBET, RHD ;
IVES, JC ;
SANFORD, PW .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1984, 31 (01) :795-800
[7]   THE EFFECTS OF CONTAMINATION ON GASEOUS X-RAY-DETECTORS [J].
RAMSEY, BD ;
ELSNER, RF ;
WEISSKOPF, MC .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1988, 270 (01) :178-184
[8]  
ROPER MD, 1992, REV SCI INS, V63
[9]   IMPROVED ENERGY RESOLUTION CAPABILITY OF AN IMAGING PROPORTIONAL COUNTER USING ELECTRON COUNTING TECHNIQUES [J].
SIEGMUND, OHW ;
CLOTHIER, S ;
CULHANE, JL ;
MASON, IM .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1983, 30 (01) :350-354
[10]  
SMITH AB, UNPUB