IMPROVEMENTS IN THE METHOD DEVELOPED FOR PERFORMING ISOELECTRIC-FOCUSING IN UNCOATED CAPILLARIES

被引:41
作者
MAZZEO, JR [1 ]
KRULL, IS [1 ]
机构
[1] NORTHEASTERN UNIV,BARNETT INST,DEPT CHEM,341 MU,BOSTON,MA 02115
来源
JOURNAL OF CHROMATOGRAPHY | 1992年 / 606卷 / 02期
关键词
D O I
10.1016/0021-9673(92)87041-6
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Capillary isoelectric focusing run times were reduced from 30 to 5 minutes through reversing the polarity and shortening the separation distance. Complete resolution was only obtained by increasing the concentration of tetramethylethylenediamine in the sample solution. Problems previously seen with acidic proteins were shown to be due to anodic drift in the ampholyte gradient, and were minimized by increasing the concentration of phosphoric acid in the anode buffer. The inability of the method to tolerate protein samples with salt is also discussed.
引用
收藏
页码:291 / 296
页数:6
相关论文
共 15 条
[1]   STEADY-STATE ELECTROLYSIS AND ISOELECTRIC-FOCUSING [J].
HAGEDORN, R ;
FUHR, G .
ELECTROPHORESIS, 1990, 11 (04) :281-289
[2]   CARRIER-FREE ZONE ELECTROPHORESIS, DISPLACEMENT ELECTROPHORESIS AND ISOELECTRIC-FOCUSING IN A HIGH-PERFORMANCE ELECTROPHORESIS APPARATUS [J].
HJERTEN, S ;
ELENBRING, K ;
KILAR, F ;
LIAO, JL ;
CHEN, AJC ;
SIEBERT, CJ ;
ZHU, MD .
JOURNAL OF CHROMATOGRAPHY, 1987, 403 :47-61
[3]   ADAPTATION OF THE EQUIPMENT FOR HIGH-PERFORMANCE ELECTROPHORESIS TO ISOELECTRIC-FOCUSING [J].
HJERTEN, S ;
ZHU, MD .
JOURNAL OF CHROMATOGRAPHY, 1985, 346 (OCT) :265-270
[4]   THEORETICAL AND EXPERIMENTAL-STUDY OF HIGH-PERFORMANCE ELECTROPHORETIC MOBILIZATION OF ISOELECTRICALLY FOCUSED PROTEIN ZONES [J].
HJERTEN, S ;
LIAO, JL ;
YAO, K .
JOURNAL OF CHROMATOGRAPHY, 1987, 387 :127-138
[5]  
HJERTEN S, 1985, J CHROMATOGR, V347, P91
[6]   FAST AND HIGH-RESOLUTION ANALYSIS OF HUMAN-SERUM TRANSFERRIN BY HIGH-PERFORMANCE ISOELECTRIC-FOCUSING IN CAPILLARIES [J].
KILAR, F ;
HJERTEN, S .
ELECTROPHORESIS, 1989, 10 (01) :23-29
[7]   SEPARATION OF THE HUMAN TRANSFERRIN ISOFORMS BY CARRIER-FREE HIGH-PERFORMANCE ZONE ELECTROPHORESIS AND ISOELECTRIC-FOCUSING [J].
KILAR, F ;
HJERTEN, S .
JOURNAL OF CHROMATOGRAPHY, 1989, 480 :351-357
[8]   CAPILLARY ISOELECTRIC-FOCUSING OF PROTEINS IN UNCOATED FUSED-SILICA CAPILLARIES USING POLYMERIC ADDITIVES [J].
MAZZEO, JR ;
KRULL, IS .
ANALYTICAL CHEMISTRY, 1991, 63 (24) :2852-2857
[9]   EXAMINATION OF VARIABLES AFFECTING THE PERFORMANCE OF ISOELECTRIC-FOCUSING IN UNCOATED CAPILLARIES [J].
MAZZEO, JR ;
KRULL, IS .
JOURNAL OF MICROCOLUMN SEPARATIONS, 1992, 4 (01) :29-33
[10]   AN EXPLANATION FOR THE PLATEAU PHENOMENON IN ISOELECTRIC-FOCUSING [J].
MOSHER, RA ;
THORMANN, W ;
BIER, M .
JOURNAL OF CHROMATOGRAPHY, 1986, 351 (01) :31-38