IMPROVED THERMOCOUPLE PSYCHROMETERS FOR MEASUREMENT OF SOIL-WATER POTENTIAL IN A TEMPERATURE-GRADIENT

被引:14
作者
CAMPBELL, GS
机构
[1] Univ. of Nottingham, School of Agriculture, Loughborough
来源
JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS | 1979年 / 12卷 / 08期
关键词
D O I
10.1088/0022-3735/12/8/018
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The improved psychrometer is constructed from high-thermal-conductivity materials to minimise temperature gradients within it. To minimise errors from condensation on the psychrometer, it is symmetrically constructed with respect to the porous windows where moisture exchange with the soil occurs. Zero offset of the new units is reduced by more than an order of magnitude compared with previous versions and measurement errors due to temperature gradients are less than one-third the value for the best designs previously available.
引用
收藏
页码:739 / 743
页数:5
相关论文
共 16 条