COMPARATIVE STUDIES OF HEMOPERFUSION DEVICES .1. INVITRO CLEARANCE CHARACTERISTICS

被引:16
作者
COONEY, DO
INFANTOLINO, W
KANE, R
机构
[1] Dept. Chem. Engin., Clarkson Coll. Technol., Potsdam, N.Y. 13676, United States
来源
BIOMATERIALS MEDICAL DEVICES AND ARTIFICIAL ORGANS | 1978年 / 6卷 / 03期
关键词
DRUG UPTAKE RATE - HEMOPERFUSION DEVICES;
D O I
10.3109/10731197809118701
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Drug adsorption studies were carried out using three charcoal-based and one resin-based hemoperfusion devices. Clearance versus time tests of up to six hours duration were performed using solutions of sodium salicylate, sodium barbital, and creatinine in a pH 7. 4 buffer which were pumped through the columns at flow rates of either 100 or 200 ml/minute. Short term clearance tests were performed at various other flow rates to determine overall mass values for the individual fluid-phase and solid-phase resistances. Studies were also done using sodium salicylate in bovine blood. The results provide information on typical adsorption capacities and drug removal rates for the four devices, and indicate the relative extents to which the solid and fluid phases control the drug uptake rates.
引用
收藏
页码:199 / 213
页数:15
相关论文
共 5 条
[1]  
Andrade J D, 1972, Proc Eur Dial Transplant Assoc, V9, P290
[2]  
DUNLOP EH, 1976, MED BIOL ENG, V14, P220, DOI 10.1007/BF02478752
[3]   RAPID DETERMINATION OF SALICYLATE IN BIOLOGICAL FLUIDS [J].
TRINDER, P .
BIOCHEMICAL JOURNAL, 1954, 57 (02) :301-303
[4]  
VALE JA, 1974, P EUR SOC ARTIF ORGA, V1, P94
[5]  
WALKER JM, 1976, KIDNEY INT, V10, pS320