TEST OF 20 SIMILAR INTENSIVE-CARE VENTILATORS IN DAILY USE CONDITIONS - EVALUATION OF ACCURACY AND PERFORMANCES

被引:7
作者
BEYDON, L
LIU, N
HASSAPOPOULOS, J
RUA, F
BONNET, F
HARF, A
DUVALDESTIN, P
机构
[1] HOP HENRI MONDOR,SERV ANESTHESIE REANIMAT,F-94010 CRETEIL,FRANCE
[2] HOP HENRI MONDOR,INSERM,U296,F-94010 CRETEIL,FRANCE
关键词
ARTIFICIAL VENTILATION EQUIPMENT; CRITICAL CARE;
D O I
10.1007/BF01706423
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Infrequent control, aging of components, may compromise the accuracy of ICU ventilators. In order to assess the reliability of ventilators during their clinical use, we bench tested a group of 20 CPU1 ventilators (Ohmeda) sampled at random in several ICU units. We found major leaks in 5 ventilators, attributable to the disposable tubings used in these systems. Mean error in expired tidal volume and corresponding standard deviation (precision) were greater than 100 ml in two. Positive end expiratory pressure measurement comprised a mean error higher than 2 cm H2O in 40% of the ventilators tested. The valve opening pressure threshold was correlated to the inspiratory flow (r = 0.81) contrary to the valve opening delay (average 138 +/- 40 ms). These two parameters did not correlate with the age of the ventilator. Our study addresses the need for periodic control of ventilator performance in order to minimize the risks of errors and malfunctions.
引用
收藏
页码:32 / 37
页数:6
相关论文
共 16 条
[1]   COMPRESSION VOLUME DURING MECHANICAL VENTILATION - COMPARISON OF VENTILATORS AND TUBING CIRCUITS [J].
BARTEL, LP ;
BAZIK, JR ;
POWNER, DJ .
CRITICAL CARE MEDICINE, 1985, 13 (10) :851-854
[2]   INSPIRATORY WORK OF BREATHING DURING SPONTANEOUS VENTILATION USING DEMAND VALVES AND CONTINUOUS-FLOW SYSTEMS [J].
BEYDON, L ;
CHASSE, M ;
HARF, A ;
LEMAIRE, F .
AMERICAN REVIEW OF RESPIRATORY DISEASE, 1988, 138 (02) :300-304
[3]  
CARA M, 1984, Agressologie, V25, P93
[4]   DEMAND AND CONTINUOUS-FLOW INTERMITTENT MANDATORY VENTILATION SYSTEMS [J].
CHRISTOPHER, KL ;
NEFF, TA ;
BOWMAN, JL ;
EBERLE, DJ ;
IRVIN, CG ;
GOOD, JT .
CHEST, 1985, 87 (05) :625-630
[5]  
CLERO B, 1984, Agressologie, V25, P129
[6]   INVESTIGATION OF THE SPONTANEOUS MODES OF BREATHING OF DIFFERENT VENTILATORS [J].
COX, D ;
TINLOI, SF ;
FARRIMOND, JG .
INTENSIVE CARE MEDICINE, 1988, 14 (05) :532-537
[7]   COMPARISON OF WORK OF BREATHING ON HIGH GAS-FLOW AND DEMAND VALVE CONTINUOUS POSITIVE AIRWAY PRESSURE SYSTEMS [J].
GIBNEY, RTN ;
WILSON, RS ;
PONTOPPIDAN, H .
CHEST, 1982, 82 (06) :692-695
[8]   INSPIRATORY WORK AND AIRWAY PRESSURE WITH CONTINUOUS POSITIVE AIRWAY PRESSURE DELIVERY SYSTEMS [J].
KATZ, JA ;
KRAEMER, RW ;
GJERDE, GE .
CHEST, 1985, 88 (04) :519-526
[9]   INSPIRATORY WORK AND RESPONSE-TIMES OF A MODIFIED PEDIATRIC VOLUME VENTILATOR DURING SYNCHRONIZED INTERMITTENT MANDATORY VENTILATION AND PRESSURE SUPPORT VENTILATION [J].
MARTIN, LD ;
RAFFERTY, JF ;
WETZEL, RC ;
GIOIA, FR .
ANESTHESIOLOGY, 1989, 71 (06) :977-981
[10]   BIAS AND PRECISION OF PULSE OXIMETERS AND ARTERIAL OXIMETERS [J].
NICKERSON, BG ;
SARKISIAN, C ;
TREMPER, K .
CHEST, 1988, 93 (03) :515-517