COMPARISON BETWEEN THE TRANS-THORACIC ELECTRICAL-IMPEDANCE METHOD AND THE DIRECT FICK AND THE DYE DILUTION METHODS FOR CARDIAC-OUTPUT MEASUREMENTS IN MAN

被引:17
作者
ENGHOFF, E [1 ]
LOVHEIM, O [1 ]
机构
[1] UNIV UPPSALA HOSP,DEPT CLIN PHYSIOL,S-75014 UPPSALA,SWEDEN
关键词
Cardiac disease; Flow measurement; Impedance cardiography; Non-invasive; Stroke volume;
D O I
10.3109/00365517909108837
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
In fourteen patients referred for preoperative evaluation of heart disease we compared the transthoracic electrical impedance method to the direct Fick (four patients) and dye dilution methods (ten patients) for estimation of cardiac output. On statistical analysis of paired differences we found no reliable agreement between the absolute impedance values and the reference methods. The reproducibility was, however, acceptable, and the coefficient of variation for impedance cardiac output was 7.8% and for impedance stroke volume 7.6% Our conclusion is that in patients with advanced heart disease cardiac output cannot be determined with accuracy with the impedance cardiographic method. In clinical practice this must be considered so important, that we hesitate to use the method for monitoring cardiac output in critically ill patients even if the reproducibility for individual changes is acceptable. © 1979 Informa UK Ltd All rights reserved: reproduction in whole or part not permitted.
引用
收藏
页码:585 / 590
页数:6
相关论文
共 19 条
[1]  
BAKER L E, 1971, Cardiovascular Research Center Bulletin (Houston), V9, P135
[2]   IMPEDANCE CARDIOGRAPHY, A METHOD TO EVALUATE QUANTITATIVELY CARDIAC-OUTPUT - COMPARISON WITH FICK PRINCIPLE [J].
BETZ, R ;
BASTANIER, CK ;
MOCELLIN, R .
BASIC RESEARCH IN CARDIOLOGY, 1977, 72 (01) :46-56
[3]  
BISHOP JM, 1963, CLIN SCI, V24, P391
[4]   MEASUREMENT OF CARDIAC-OUTPUT BY ELECTRICAL IMPEDANCE AT REST AND DURING EXERCISE [J].
DENNISTON, JC ;
MAHER, JT ;
REEVES, JT ;
CRUZ, JC ;
CYMERMAN, A ;
GROVER, RF .
JOURNAL OF APPLIED PHYSIOLOGY, 1976, 40 (01) :91-95
[5]  
ENGHOFF E, 1970, ACTA SOC MED UPS SCI, V75, P157
[6]  
ENGHOFF E, 1972, ACTA MED SCAND S538
[7]   MEASUREMENT OF CARDIAC-OUTPUT BY IMPEDANCE CARDIOGRAPHY IN PATIENTS WITH MYOCARDIAL-INFARCTION - COMPARATIVE EVALUATION OF IMPEDANCE AND DYE DILUTION METHODS [J].
GABRIEL, S ;
ATTERHOG, JH ;
ORO, L ;
EKELUND, LG .
SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 1976, 36 (01) :29-34
[8]   SPECIFIC RESISTANCE OF CANINE BLOOD AT BODY-TEMPERATURE [J].
GEDDES, LA ;
DACOSTA, CP .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1973, BM20 (01) :51-53
[9]  
Grogler F M, 1976, Thoraxchir Vask Chir, V24, P291
[10]  
HANDT A, 1977, CLIN NEPHROL, V7, P61