A MULTICENTER EVALUATION OF A SINGLE-PASS LEAD VDD PACING SYSTEM

被引:54
作者
CURZIO, G [1 ]
机构
[1] MONALDI HOSP,DEPT ELECTROSTIMULAT,NAPLES,ITALY
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1991年 / 14卷 / 03期
关键词
DIFFERENTIAL BIPOLAR SENSING; SINGLE-PASS LEAD; ATRIAL SYNCHRONY;
D O I
10.1111/j.1540-8159.1991.tb04092.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Since June 1985 until April 1989, 237 patients (130 males, 107 females, aged 22 to 95 years, mean 71) with symptomatic AV conduction disturbances and competent sinus node, were implanted with a single-pass lead VDD pacing system in 30 centers and followed-up for at least 6 months. The ventricular pacing lead incorporated two atrial ring 3-cm apart electrodes, positioned within the right atrial cavity without contact with the heart wall, in order to detect the atrial activity, which is differentially processed by the pacemaker. At implant, mean atrial electrogram amplitude, derived from a custom pacemaker system analyzer (PSA) with the same input filter of the pacemaker was 1.7 +/- 0.8 mV (n = 93). In all cases, atrial sensitivity at implant was the default value +/- 0.15 mV. The atrial tracking capability of the pacing systems was assessed within the month and every 6 months after implantation by means of clinical evaluation, resting ECG, 24-hour Holter monitoring and the following tests: exercise stres testing, mental stress, isometric exercise, and nifedipine test. These tests evoke an increase of atrial rate in consequence of metabolic needs or as a reflex response. The criterion used to evaluate the correct operation of the system was the percentage of atrial synchronization. This was defined as the ratio between atrial triggered ventricular paced complexes and all ventricular paced complexes. All monitorings showed a ratio higher than or equal to 98% in a percentage of patients not lower than 95%. Mean follow-up was 385 days (range 183-1,370 days). During follow-up, atrial sensitivity was never reprogrammed and inappropriate triggered pacing due to myopotential activity was never noted in any tests.
引用
收藏
页码:434 / 442
页数:9
相关论文
共 20 条
[1]  
ALBUNNI M, 1987, CARDIOSTIMOLAZIONE, V5, P183
[2]   SENSING CHARACTERISTICS OF UNIPOLAR AND BIPOLAR ORTHOGONAL FLOATING ATRIAL ELECTRODES - MORPHOLOGY AND SPECTRAL-ANALYSIS [J].
AUBERT, AE ;
ECTOR, H ;
DENYS, BG ;
DEGEEST, H .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1986, 9 (03) :343-359
[3]  
BORASTEROS C, 1987, PACE, V10, P648
[4]  
CURZIO G, 1987, PACE, V10, P662
[5]  
CURZIO G, 1988, CARDIOSTIMOLAZIONE, V6, P257
[6]  
CURZIO G, 1985, CARDIAC PACING ELECT, P865
[7]   SINUS NODE DYSFUNCTION [J].
DHINGRA, RC .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1983, 6 (05) :1062-1069
[8]  
DURZIO G, 1986, PROGR CLIN PACING, P366
[9]   A TIME-RELATED STUDY OF THE HEMODYNAMIC BENEFIT OF ATRIOVENTRICULAR SYNCHRONOUS PACING EVALUATED BY DOPPLER ECHOCARDIOGRAPHY [J].
FAERESTRAND, S ;
OHM, OJ .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1985, 8 (06) :838-848
[10]   COMPARISON OF RESTING HEMODYNAMIC INDEXES AND EXERCISE PERFORMANCE DURING ATRIAL SYNCHRONIZED AND ASYNCHRONOUS VENTRICULAR PACING [J].
FANANAPAZIR, L ;
SRINIVAS, V ;
BENNETT, DH .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1983, 6 (02) :202-209