NEW PROGRAMMABLE TIMER DESIGNED FOR PULSED NMR

被引:6
作者
CARON, F
HERZOG, RF
机构
[1] EUROPEAN MOLEC BIOL LAB,D-6900 HEIDELBERG,FED REP GER
[2] CNRS,CTR GENET MOLEC,F-91190 GIF SUR YVETTE,FRANCE
关键词
D O I
10.1016/S0022-2364(78)80001-8
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
This article describes a sequencing circuit which generates accurate time intervals for use in pulsed Fourier transform NMR spectrometers. This microprogrammable unit allows control of the spectrometer and definition of the next synchronization event for each time interval. The time intervals range from 1 μsec up to days with basic 12-bit accuracy and a 10-MHz clock. Subsequences of time intervals can be repeated a given number of times. The implementation presented here is directly interfaced to a NOVA 800 computer; interfacing to the CAMAC system or to other computers is straightforward. © 1978 Academic Press, Inc. All rights reserved.
引用
收藏
页码:357 / 362
页数:6
相关论文
共 9 条
[1]  
Ellett J.D., 1971, ADV MAGN RESON, V662, P117
[2]   FOURIER TRANSFORM STUDY OF NMR SPIN-LATTICE RELAXATION BY PROGRESSIVE-SATURATION [J].
FREEMAN, R ;
HILL, HDW .
JOURNAL OF CHEMICAL PHYSICS, 1971, 54 (08) :3367-&
[3]  
GUERON M, 1977, RAPPORTS INTERNES GR, V3
[4]  
MICHAUD JP, 1976, RAPPORT INTERNE GROU, V1
[5]  
REDFIELD AG, 1971, ADV MAGN RESON, V5, P81
[6]  
RAPPORTS INTERNES GR, V6
[7]  
1972, HOW USE NOVA COMPUTE
[8]  
RAPPORTS INTERNES GR, V5
[9]  
1976, 67110 MON MEM INC MI