LATE TRANSCATHETER REMOVAL OF A RASHKIND PDA OCCLUSION DEVICE FOR PERSISTENT HEMOLYSIS USING A MODIFIED TRANSSEPTAL SHEATH

被引:10
作者
GRIFKA, RG
OLAUGHLIN, MP
MULLINS, CE
机构
[1] TEXAS CHILDRENS HOSP,6621 FANNIN,HOUSTON,TX 77030
[2] BAYLOR COLL MED,HOUSTON,TX 77030
来源
CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS | 1992年 / 25卷 / 02期
关键词
PATENT DUCTUS ARTERIOSUS; RETRIEVAL; REOCCLUSION; TRANSSEPTAL SHEATH;
D O I
10.1002/ccd.1810250210
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
A patient developed hemolysis and hematuria 2 d after placement of a 12 mm Rashkind PDA occlusion device. Six days after placement, the device was snared from the ductus arteriosus and removed through a modified transseptal sheath. A 17 mm PDA occlusion device was successfully placed in the PDA, with resolution of the hematuria.
引用
收藏
页码:140 / 143
页数:4
相关论文
共 6 条
[1]   CATHETER OCCLUSION OF THE PERSISTENTLY PATENT DUCTUS-ARTERIOSUS [J].
DYCK, JD ;
BENSON, LN ;
SMALLHORN, JF ;
MCLAUGHLIN, PR ;
FREEDOM, RM ;
ROWE, RD .
AMERICAN JOURNAL OF CARDIOLOGY, 1988, 62 (16) :1089-1092
[2]  
GREGORIC I, 1990, CIRCULATION S3, V82, P582
[3]  
LADUSANS EJ, 1989, BRIT HEART J, V61, P548
[4]   TRANS-SEPTAL LEFT HEART CATHETERIZATION - EXPERIENCE WITH A NEW TECHNIQUE IN 520 PEDIATRIC AND ADULT PATIENTS [J].
MULLINS, CE .
PEDIATRIC CARDIOLOGY, 1983, 4 (03) :239-245
[5]  
OLAUGHLIN MP, 1990, CIRCULATION, V82, P582
[6]   NONSURGICAL CLOSURE OF PATENT DUCTUS-ARTERIOSUS - CLINICAL-APPLICATION OF THE RASHKIND PDA OCCLUDER SYSTEM [J].
RASHKIND, WJ ;
MULLINS, CE ;
HELLENBRAND, WE ;
TAIT, MA .
CIRCULATION, 1987, 75 (03) :583-592