APPLICATION OF GAAS AND CDTE PHOTOCONDUCTOR DETECTORS TO X-RAY FLASH RADIOGRAPHY

被引:7
作者
MATHY, F [1 ]
CUZIN, M [1 ]
GAGELIN, JJ [1 ]
MERMET, R [1 ]
PIAGET, B [1 ]
RUSTIQUE, J [1 ]
VERGER, L [1 ]
HAUDUCOEUR, A [1 ]
NICOLAS, P [1 ]
LEDAIN, L [1 ]
HYVERNAGE, M [1 ]
机构
[1] CEA,DIRECT APPLICAT MIL,CE VAUJOURS MORONVILLIERS,F-77181 COURTRY,FRANCE
关键词
D O I
10.1016/0168-9002(92)91240-A
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Some insulating GaAs and CdTe:Cl photoconductor probes were qualified on high energy X-ray single-shot flash generators. The estimated minimum detected dose per flash corresponding to a 230 mrad direct beam attenuated by 200 mm lead was 20 mu rad. The dynamic range was about 4 decades in amplitude or charge, with a good linearity. Such detectors, by locating the origin of the parasitic scattered beam, could be used to eliminate this parasitic beam in X-ray flash radiography in detonics experiments. Imaging possibilities are mentioned, as well as X-ray generator monitoring with such detectors or with neutron preirradiated photoconductors.
引用
收藏
页码:615 / 618
页数:4
相关论文
共 5 条
[2]   CDTE IN PHOTOCONDUCTIVE APPLICATIONS FAST DETECTOR FOR METROLOGY AND X-RAY-IMAGING [J].
CUZIN, M .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1992, 322 (03) :341-351
[3]  
GLASSER F, 1990, LETIDSYSSESA90327FGM
[4]  
MOULIN H, 1989, THESIS GRENOBLE 1
[5]   ELECTRONIC-PROPERTIES OF CHLORINE DOPED CADMIUM TELLURIDE USED AS HIGH-ENERGY PHOTOCONDUCTIVE DETECTOR [J].
VERGER, L ;
CUZIN, M ;
GAUDE, G ;
GLASSER, F ;
MATHY, F ;
RUSTIQUE, J ;
SCHAUB, B .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1992, 322 (03) :357-362