POROUS POLYURETHANE VASCULAR PROSTHESES WITH VARIABLE COMPLIANCES

被引:39
作者
LIU, SQ
KODAMA, M
机构
[1] Biomaterial Laboratory, Research Institute for Polymers and Textiles, Tsukuba, Ibaraki, 305
来源
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH | 1992年 / 26卷 / 11期
关键词
D O I
10.1002/jbm.820261108
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A new technique for the preparation of porous vascular prostheses was investigated. Polyurethane solution (5 to 15 wt%) was injected into a mold. After freezing at low temperature (0-degrees-C - 196-degrees-C), solvents were dissolved out with water at 0-degrees-C to form porous tubes. The average pore size (several microns to 70 mum), pore occupation (10% to 51%), and compliance (3% to 35%) were easily changed by changing polyurethane concentration, freezing temperature, and freezing methods. Compliances could be decreased gradually by heat treatment. This technique can give a proper pore size (30-60 mum) for tissue ingrowth, and a suitable compliance for matching with arteries and veins. This method might give a desired compliant graft for artificial implantation with the presently valid medical polymers.
引用
收藏
页码:1489 / 1502
页数:14
相关论文
共 18 条
[1]  
ANNIS D, 1978, T AM SOC ART INT ORG, V24, P209
[2]  
CALLOW AD, 1988, INT ANGIOL, V7, P246
[3]   SOLVENT-STIMULATED SURFACE REARRANGEMENT OF POLYURETHANES [J].
CHEN, JH ;
RUCKENSTEIN, E .
JOURNAL OF COLLOID AND INTERFACE SCIENCE, 1990, 135 (02) :496-507
[4]  
CLARK RE, 1976, SURG FORUM, V27, P208
[5]  
CLOWES AW, 1986, AM J PATHOL, V123, P220
[6]  
CRONENWETT JL, 1982, BIOL SYNTHETIC VASCU, P595
[7]  
DILLEY RJ, 1988, ARCH SURG-CHICAGO, V123, P691
[8]  
HAYASHI K, 1989, J BIOMED MATER RES-A, V23, P229
[9]  
KARDOS J L, 1974, Biomaterials Medical Devices and Artificial Organs, V2, P387
[10]  
KINLEY CE, 1974, CAN J SURG, V17, P74