CALIBRATION AND ANALYSIS OF X HOT-WIRE PROBE SIGNALS

被引:20
作者
BRUUN, HH
NABHANI, N
ALKAYIEM, HH
FARDAD, AA
KHAN, MA
HOGARTH, E
机构
[1] Dept. of Mech. and Manuf. Eng., Bradford Univ.
关键词
X Hot-Wire Probes;
D O I
10.1088/0957-0233/1/8/019
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
The authors deal with the use of X hot-wire probes for measurements of two velocity components. Using the effective velocity concept, four different analytical expressions for the yaw response of the hot wire signals have been considered. Using a set of reference calibration data the curve-fit accuracy of the four methods are first investigated. Then the related sum and difference signal analysis methods are compared and finally the velocity component errors caused by uncertainty in the value of the mean yaw angle are established.
引用
收藏
页码:782 / 785
页数:4
相关论文
共 11 条
[1]  
ALKAYIEM HH, 1989, THESIS BRADFORD U UK
[2]  
Bradshaw P., 1971, INTRO TURBULENCE ITS
[3]   VELOCITY CALIBRATION RELATIONSHIPS FOR HOT-WIRE ANEMOMETRY [J].
BRUUN, HH ;
KHAN, MA ;
ALKAYIEM, HH ;
FARDAD, AA .
JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1988, 21 (02) :225-232
[4]  
BRUUN HH, 1986, ADV HOT WIRE ANEMOME
[5]  
DAVIES POA, 1968, P S INSTRUM DATA P T
[6]  
FARDAD AA, 1989, THESIS BRADFORD U UK
[7]  
FRIEHE CA, 1968, J APPL MECH, V90, P655
[8]  
Hinze J, 1959, TURBULENCE
[9]  
JORGENSEN FE, 1971, DISA INFORM, V11, P31
[10]  
KHAN MA, 1990, THESIS BRADFORD U UK