The optimization of flame atomic absorption instrumental variables such as hollow-cathode lamp current, slit width, slit height, flame type and stoichiometry, burner position, resonance line, and integration time for best precision or signal-to-noise ratio is discussed. Because different types of noise are limiting in different absorbance regions, optimum values of instrumental variables can vary with the absorbance measured. © 1979.