The Fricke ferrous sulphate dosemeter has been calibrated against a plane-parallel extrapolation chamber for 60 kvp x-rays, HVL 0.20 mm Al, apparent quantum energy 12 kev. The energy absorbed in the Fricke dosemeter was assessed by the 'relative effective density' method of Loevinger and Yaniv. A tabulation of errors, systematic and random, inherent in these measurements, has been made, end their effects evaluated. A value of GFe+++ = 14.63 ± 0.26 was obtained.
引用
收藏
页码:61 / &
相关论文
共 19 条
[1]   CAVITY IONIZATION AS A FUNCTION OF WALL MATERIAL [J].
ATTIX, FH ;
VERGNE, LD ;
RITZ, VH .
JOURNAL OF RESEARCH OF THE NATIONAL BUREAU OF STANDARDS, 1958, 60 (03) :235-243
[3]  
BOAG JW, 1966, RAD DOSIMETRY, V2, P16
[4]  
COTTIN M, 1956, J CHIM PHYS, V53, P268
[6]   DOSIMETRIE ABSOLUE POUR DES RAYONS-X DENERGIE MOYENNE 23-KEV ETALONNAGE DU SULFATE FERREUX [J].
FRILLEY, M ;
LEFORT, M .
JOURNAL DE PHYSIQUE ET LE RADIUM, 1959, 20 (05) :516-518
[7]   FERRIC ION YIELDS IN FERROUS SULFATE SOLUTIONS IRRADIATED WITH LOW-ENERGY X-RAYS [J].
GEVANTMAN, LH ;
PESTANER, JF .
JOURNAL OF CHEMICAL PHYSICS, 1959, 31 (04) :1140-1140
[8]   X-IRRADIATION AND GAMMA-IRRADIATION OF FERROUS SULFATE IN DILUTE AQUEOUS SOLUTION [J].
HAYBITTLE, JL ;
SAUNDERS, RD ;
SWALLOW, AJ .
JOURNAL OF CHEMICAL PHYSICS, 1956, 25 (06) :1213-1217
[9]  
HUBBELL JH, 1965, 8681 NAT BUR STAND R
[10]  
JENNINGS WA, 1961, BRIT J RADIOL S10