A PROJECTOR, BACK-PROJECTOR PAIR WHICH ACCOUNTS FOR THE 2-DIMENSIONAL DEPTH AND DISTANCE DEPENDENT BLURRING IN SPECT

被引:41
作者
PENNEY, BC [1 ]
KING, MA [1 ]
KNESAUREK, K [1 ]
机构
[1] UNIV ROCHESTER,MED CTR,DIV NUCL MED,ROCHESTER,NY 14642
基金
美国国家卫生研究院;
关键词
D O I
10.1109/23.106698
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The speed and accuracy of iterative reconstruction routines for single photon emission computed tomography (SPECT) are highly dependent on the projection and backprojection algorithms used. A projector-back-projector pair which accounts for the two-dimensional (2D), spatially dependent blurring encountered in SPECT and which reduces the computational load has been developed. These algorithms assume that the attenuator is a uniform elliptical cylinder that is invariant in the axial direction. They further assume that the blur can be modeled using the sum of several separable functions, e.g., separate Gaussian functions for the geometric and scatter components. When implemented on an array processor of moderate computational power, a projection or back-projection of a SPECT study, with 64 64×64 pixel projection images and a 64×64×64 reconstruction space, can be completed in less than 2.4 hours, or 2.3 minutes per transverse slice. Copyright © 1990 by The Institute of Electrical and Electronics Engineers, Inc.
引用
收藏
页码:681 / 686
页数:6
相关论文
共 22 条
[1]  
ABRAMOWITZ M, 1964, HDB MATH FUNCTIONS, P932
[2]  
ATKINS FB, 1978, THESIS U CHICAGO CHI
[3]  
AXELSSON B, 1984, J NUCL MED, V25, P490
[4]   PRINCIPLES OF COMPUTER-ASSISTED TOMOGRAPHY (CAT) IN RADIOGRAPHIC AND RADIOISOTOPIC IMAGING [J].
BROOKS, RA ;
DICHIRO, G .
PHYSICS IN MEDICINE AND BIOLOGY, 1976, 21 (05) :689-732
[5]  
Budinger T. F., 1979, Image reconstruction from projections. Implementation and applications, P147
[6]   COMPENSATION FOR COLLIMATOR DIVERGENCE IN SPECT USING INVERSE MONTE-CARLO RECONSTRUCTION [J].
FLOYD, CE ;
JASZCZAK, RJ ;
MANGLOS, SH ;
COLEMAN, RE .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1988, 35 (01) :784-787
[7]  
FLOYD CE, 1986, J NUCL MED, V27, P1577
[8]   COMPENSATION OF SPATIAL SYSTEM RESPONSE IN SPECT WITH CONJUGATE-GRADIENT RECONSTRUCTION TECHNIQUE [J].
FORMICONI, AR ;
PUPI, A ;
PASSERI, A .
PHYSICS IN MEDICINE AND BIOLOGY, 1989, 34 (01) :69-84
[9]  
FREY EC, 1989, MED PHYS, V16, P462
[10]  
GILLAND DR, 1988, J NUCL MED, V29, P643