TIMING CHARACTERISTICS OF A MICROCHANNEL PLATE AND FINE MESH PHOTOMULTIPLIER TUBES IN A 1-T FIELD

被引:26
作者
KICHIMI, H
SAGAWA, H
YOSHIMURA, Y
KISHIDA, T
TAMURA, N
SUZUKI, H
TAGUCHI, T
机构
[1] UNIV TOKYO,DEPT PHYS,BUNKYO KU,TOKYO 113,JAPAN
[2] OKAYAMA UNIV,DEPT PHYS,OKAYAMA 700,JAPAN
[3] HAMAMATSU PHOTON KK,CTR ELECTRON TUBE,IWATA,SHIZUOKA 438,JAPAN
关键词
D O I
10.1016/0168-9002(93)90390-4
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Timing characteristics of a micro-channel plate (MCP) and a fine mesh (FM) photomultiplier tubes were investigated in magnetic fields up to 1.5 T as functions of the field strength and direction. The r.m.s. values of transit time spread of the MCP and the FM tubes in 1 T field were found to be 40 ps and 140 ps, respectively. For a simulated scintillation light equivalent to 160 photoelectrons, the FM tube gave a time resolution better than 100 ps even in a 1 T field in spite of a large gain reduction to less than a 10%.
引用
收藏
页码:451 / 457
页数:7
相关论文
共 18 条
[1]  
ABE K, 1992, KEK REPORT, V923
[2]  
Atwood W.B., 1980, LECTURE SLAC SUMMER, P287
[3]   PERFORMANCE OF TRANSMISSION DYNODE PHOTOMULTIPLIERS IN MAGNETIC-FIELDS UP TO 1.5-T [J].
FINSEL, G ;
FUNK, M ;
MORDHORST, S ;
DUHM, HH ;
KRAUSE, H ;
LANGKAU, R ;
SCOBEL, W .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1990, 290 (2-3) :450-456
[4]   TESTS OF CONVENTIONAL AND MICROCHANNEL PLATE PHOTOMULTIPLIERS AS COMPONENTS OF A SCINTILLATION TIME-OF-FLIGHT SYSTEM [J].
GILES, RT ;
PIPKIN, FM ;
WOLINSKI, JP .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1986, 252 (01) :41-52
[5]  
GILES TT, 1992, CLEO2 CORN U INT REP
[6]   DECAY TIME AND LIGHT YIELD MEASUREMENTS FOR PLASTIC SCINTILLATING FIBERS [J].
HAWKES, CM ;
KUHLEN, M ;
MILLIKEN, B ;
STROYNOWSKI, R ;
WICKLUND, E ;
SHIMIZU, T ;
SHINJI, O .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1990, 292 (02) :329-336
[7]   TRANSMISSION SILICON PHOTOEMITTERS AND ELECTRON MULTIPLIERS [J].
HOWORTH, JR ;
FOLKES, JR ;
PALMER, IC .
JOURNAL OF PHYSICS D-APPLIED PHYSICS, 1976, 9 (05) :785-794
[8]  
KUHLEM M, 1990, CALT681648
[9]   RECENT DEVELOPMENT OF PHOTOMULTIPLIER TUBES FOR NUCLEAR AND MEDICAL APPLICATIONS [J].
KUME, H ;
SUZUKI, S ;
OBA, K .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1985, 32 (01) :355-359
[10]   ULTRAFAST MICROCHANNEL PLATE PHOTOMULTIPLIERS [J].
KUME, H ;
KOYAMA, K ;
NAKATSUGAWA, K ;
SUZUKI, S ;
FATLOWITZ, D .