CALCULATIONS OF ION-CHAMBER DISPLACEMENT EFFECT CORRECTIONS FOR MEDIUM-ENERGY X-RAY DOSIMETRY

被引:27
作者
MA, CM
NAHUM, AE
机构
[1] INST CANC RES,JOINT DEPT PHYS,SUTTON SM2 5PT,SURREY,ENGLAND
[2] ROYAL MARSDEN HOSP,SUTTON SM2 5PT,SURREY,ENGLAND
关键词
D O I
10.1088/0031-9155/40/1/005
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This paper investigates the displacement effect corrections for ionization chambers calibrated in air in terms of air kerma and used at a depth in a phantom irradiated by medium-energy x-ray beams for tube potentials between 100 kV and 300 kV. The EGS4 (electron gamma shower version 4) Monte Carlo code system has been used to simulate the coupled transport of photons and electrons. The chamber displacement correction factor, p(dis), has been calculated both using a simple photon attenuation and scattering method and by direct Monte Carlo calculation of the ratio of the water kerma at a depth in a phantom to the water kerma in a low-density (the same density as that of air) water cavity with its centre placed at the same depth in the phantom irradiated by medium-energy x-ray beams. Good agreement has been achieved between the p(dis) factors obtained using the simple method and those obtained by the direct Monte Carlo simulation. In contrast to the values recommended by the IAEA code of practice, which exceed unity by up to 10%, our p(dis) values for a stemless NE2571 chamber vary between 0.994 and 1.001 for the same energy range and for a stemless NE2561 chamber between 0.974 and 0.993. The beam quality correction factor and stem correction factor have been investigated in separate works. The ratios of the overall correction factors calculated in this work for an NE2571 chamber to that for an NE2561 chamber are consistent with the recent experimental results.
引用
收藏
页码:45 / 62
页数:18
相关论文
共 28 条
[1]  
[Anonymous], 1987, TECHNICAL REPORT SER, V277
[2]  
[Anonymous], 1983, MED PHYS, V10, P741
[3]  
[Anonymous], [No title captured]
[4]  
Bielajew AF, 1988, MONTE CARLO TRANSPOR, P115
[5]   CHAMBER REPLACEMENT CORRECTION IN ABSORBED DOSE CALIBRATIONS [J].
BURNS, JE .
MEDICAL PHYSICS, 1992, 19 (01) :209-210
[6]   DISPLACEMENT CORRECTIONS USED IN ABSORBED DOSE DETERMINATION [J].
CUNNINGHAM, JR ;
SONTAG, MR .
MEDICAL PHYSICS, 1980, 7 (06) :672-676
[7]  
Johansson K.-A., 1978, National and International Standardization of Radiation Dosimetry, P243
[8]   CALCULATION OF ABSORBED DOSE RATIOS USING CORRELATED MONTE-CARLO SAMPLING [J].
MA, CM ;
NAHUM, AE .
MEDICAL PHYSICS, 1993, 20 (04) :1189-1199
[9]   MONTE-CARLO CALCULATED STEM EFFECT CORRECTIONS FOR NE2561 AND NE2571 CHAMBERS IN MEDIUM-ENERGY X-RAY-BEAMS [J].
MA, CM ;
NAHUM, AE .
PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (01) :63-72
[10]   IMPLEMENTATION OF A MONTE-CARLO CODE ON A PARALLEL COMPUTER-SYSTEM [J].
MA, CM .
PARALLEL COMPUTING, 1994, 20 (07) :991-1005