LACK OF AGREEMENT BETWEEN MEASUREMENT OF EJECTION FRACTION BY IMPEDANCE CARDIOGRAPHY VERSUS RADIONUCLIDE VENTRICULOGRAPHY

被引:12
作者
BOWLING, LS
SAGEMAN, WS
OCONNOR, SM
COLE, R
AMUNDSON, DE
机构
[1] USN HOSP,DEPT INTERNAL MED,DIV CARDIOL,SAN DIEGO,CA 92134
[2] USN HOSP,DEPT CRIT CARE MED,SAN DIEGO,CA 92134
[3] USN HOSP,DEPT RADIOL & CLIN INVEST,DIV NUCL MED,SAN DIEGO,CA 92134
关键词
CARDIOGRAPHY; IMPEDANCE; THORACIC; EJECTION FRACTION; LEFT VENTRICULAR FUNCTION; SYSTOLIC TIME INTERVALS; RADIONUCLIDE VENTRICULOGRAPHY; CARDIAC EMERGENCIES; HEART; CRITICAL ILLNESS;
D O I
10.1097/00003246-199310000-00021
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective: To determine the limits of agreement between left ventricular ejection fraction estimated using systolic time intervals from impedance cardiography and left ventricular ejection fraction estimated by radionuclide ventriculography. Design: A prospective study for diagnostic tests using radionuclide ventriculography as the criterion standard. Setting: A large, military teaching hospital. Patients: Twenty ambulatory adults scheduled for radionuclide ventriculography. Measurements and Main Results: A regression equation to estimate ejection fraction from systolic time intervals is available in a widely used impedance-based cardiac monitoring device. The estimated ejection fraction is then used in an equation with stroke volume estimated by the same device to calculate an end-diastolic volume. We studied the agreement of the ejection fraction as estimated by this device with the ejection fraction estimated by radionuclide ventriculography by obtaining simultaneous estimates of ejection fraction over a broad range of adult patients. Twenty ejection fraction pairs were analyzed. The correlation of ejection fraction by impedance cardiography to ejection fraction by radionuclide ventriculography was significant (r2 = .55; p < .002). However, the mean difference between the technologies was -8.85%, with a standard deviation of the differences of 7.15%, resulting in a 95% confidence range for agreement of -23.2% to +5.5%. Conclusions: The 95% confidence range defining the limits of agreement between ejection fraction by impedance cardiography and ejection fraction by radionuclide ventriculography is not clinically acceptable. In the opinion of the authors impedance cardiography should not be used in place of radionuclide ventriculography for the assessment of ejection fraction at this time.
引用
收藏
页码:1523 / 1527
页数:5
相关论文
共 9 条
  • [1] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [2] ARE EJECTION FRACTIONS BASED ON REGRESSION EQUATIONS FOR SYSTOLIC-TIME RATIOS RELIABLE
    BOWLING, LS
    AMUNDSON, DE
    [J]. CRITICAL CARE MEDICINE, 1992, 20 (10) : 1459 - 1460
  • [3] BRAUNWALD E, 1992, HEART DISEASE TXB CA, P377
  • [4] ANALYSIS OF LEFT-VENTRICULAR FUNCTION FROM MULTIPLE GATED ACQUISITION CARDIAC BLOOD POOL IMAGING - COMPARISON TO CONTRAST ANGIOGRAPHY
    BUROW, RD
    STRAUSS, HW
    SINGLETON, R
    POND, M
    REHN, T
    BAILEY, IK
    GRIFFITH, LC
    NICKOLOFF, E
    PITT, B
    [J]. CIRCULATION, 1977, 56 (06) : 1024 - 1028
  • [5] MEASUREMENT OF EJECTION FRACTION BY BIOIMPEDANCE METHOD
    CAPAN, LM
    BERNSTEIN, DP
    PATEL, KP
    SANGER, J
    TURNDORF, H
    [J]. CRITICAL CARE MEDICINE, 1987, 15 (04) : 402 - 402
  • [6] RELATIONSHIP OF ALTERATIONS IN SYSTOLIC TIME INTERVALS TO EJECTION FRACTION IN PATIENTS WITH CARDIAC DISEASE
    GARRARD, CL
    WEISSLER, AM
    DODGE, HT
    [J]. CIRCULATION, 1970, 42 (03) : 455 - &
  • [7] GIBBONS RJ, 1991, CARDIAC IMAGING, P1029
  • [8] BEDSIDE TECHNICS FOR EVALUATION OF VENTRICULAR FUNCTION IN MAN
    WEISSLER, AM
    HARRIS, WS
    SCHOENFELD, CD
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1969, 23 (04) : 577 - +
  • [9] 1989, CARDIODYNAMIC DATA P