ION CHROMATOGRAPHY-INDUCTIVELY-COUPLED PLASMA-ATOMIC EMISSION-SPECTROMETRY (IC-ICP-AES) AS A METHOD FOR DETERMINING TRACE-METALS IN ESTUARINE WATER

被引:10
作者
COFFEY, MJ [1 ]
JICKELLS, TD [1 ]
机构
[1] UNIV E ANGLIA,DEPT ENVIRONM SCI,NORWICH NR4 7TJ,NORFOLK,ENGLAND
关键词
TRACE METALS; ESTUARIES; SPECTROMETRY; ION-EXCHANGE RESINS; HUMBER ESTUARY;
D O I
10.1006/ecss.1995.0026
中图分类号
Q17 [水生生物学];
学科分类号
071004 ;
摘要
A method by which estuarine water samples have been successfully analysed for the dissolved trace metals Mn, Fe, Cu and Zn using an ion chromatograph (IC) coupled directly to an inductively coupled plasma-atomic emission spectrometry (ICP-AES) system is described. Direct determination by ICP-AES of these metals in the samples was not possible since detection limits of this instrument are too high in the estuarine water matrix. The coupling of the IC to this ICP-AES system lowers detection limits sufficiently to allow analysis in two ways. First, the sample is pre-concentrated on-line; second, much of the background emission caused by the matrix is removed prior to analysis, which improves the signal-to-noise ratio, This latter feature of the method is shown to remove observable matrix effects from sample analysis, precluding the need for the use of the method of standard additions, and allowing samples from the entire salinity gradient to be accurately measured against an aqueous standard calibration line. The only required pre-treatment for samples is filtration (0.2 mu m) and buffering to pH 5.4 with ammonium acetate; multi-elemental analysis is complete within 10 min. For the metals determined, the limits of determination were Cu:1.11, Fe:1.46, Mn:0.20, Zn:0.75 mu g l(-1) (17, 26, 4 and 11 nM, respectively). Precision on saline samples was estimated at Cu:1.41, Fe:1.23, Mn:0.27, Zn:1.11 mu g l(-1) (22, 22, 5 and 17 nM, respectively), which improved for freshwater samples. The method has a large linear range and is accurate based on analyses of standard reference material.
引用
收藏
页码:379 / 386
页数:8
相关论文
共 11 条
[1]   EXAMINATION OF THE RANGE OF COPPER COMPLEXING LIGANDS IN NATURAL-WATERS USING A COMBINATION OF CATHODIC STRIPPING VOLTAMMETRY AND COMPUTER-SIMULATION [J].
APTE, SC ;
GARDNER, MJ ;
RAVENSCROFT, JE ;
TURRELL, JA .
ANALYTICA CHIMICA ACTA, 1990, 235 (02) :287-297
[2]   AN INTERCALIBRATION FOR TRACE-METALS IN SEAWATER [J].
BEWERS, JM ;
DALZIEL, J ;
YEATS, PA ;
BARRON, JL .
MARINE CHEMISTRY, 1981, 10 (03) :173-193
[4]   DIRECT DETERMINATION OF DISSOLVED BARIUM IN SEA-WATER BY INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROMETRY [J].
DEHAIRS, F ;
NEYBERGH, H ;
HOENIG, M .
ANALYTICA CHIMICA ACTA, 1989, 222 (01) :55-61
[5]  
JICKELLS T, 1992, ANAL PROC, V29, P288
[7]  
RIVIELLO JM, 1992, LC GC INT, V4, P25
[8]   CHELATION ION CHROMATOGRAPHY AS A METHOD FOR TRACE ELEMENTAL ANALYSIS IN COMPLEX ENVIRONMENTAL AND BIOLOGICAL SAMPLES [J].
SIRIRAKS, A ;
KINGSTON, HM ;
RIVIELLO, JM .
ANALYTICAL CHEMISTRY, 1990, 62 (11) :1185-1193
[10]  
THOMPSON M, 1989, HDB INDUCTIVELY COUP, P201