COULOMETRIC ASSAY OF PRIMARY STANDARDS POTASSIUM DICHROMATE AND AMMONIUM HEXANITRATOCERATE

被引:31
作者
KNOECK, J
DIEHL, H
机构
[1] Department of Chemistry, Iowa State University, Ames
关键词
D O I
10.1016/0039-9140(69)80269-9
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The high-precision coulometric titration apparatus of Eckfeldt and Shaffer (Leeds & Northrup Company) has been used for the precision assay of two primary standard materials for oxidation-reduction work. The purity found for NBS 136b Potassium Dichromate. 99.975 %, standard deviation 0.002 %, checks the earlier value of Marinenko and Taylor. Ammonium hexanitratocerate, if low in thorium, is also an excellent primary standard, two recent commercial preparations having purities of 99.972 and 99.984%, a higher standard deviation, 0.005 %, reflecting a minor difficulty in end-point detection caused by platinum oxide formation on the indicating electrodes. © 1969.