SIGNIFICANCE OF PHYSICAL PARAMETERS FOR THE EFFICACY OF RADIOFREQUENCY CATHETER ABLATION

被引:1
作者
HOFFMANN, E [1 ]
DORWARTH, U [1 ]
PULTER, R [1 ]
GOKEL, M [1 ]
STEINBECK, G [1 ]
机构
[1] UNIV MUNICH,KLINIKUM GROSSHADERN,INST PATHOL,W-8000 MUNICH 70,GERMANY
来源
BIOMEDIZINISCHE TECHNIK | 1992年 / 37卷 / 04期
关键词
RADIOFREQUENCY ENERGY; CATHETER ABLATION; PHASE ANGLE; MYOCARDIAL NECROSIS;
D O I
10.1515/bmte.1992.37.4.62
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Radiofrequency catheter ablation has been shown to be an effective form of accessory pathways in patients with WPW-syndrome and other supraventricular tachycardias. However, the biophysical parameters so far used in vivo neither correlated with the size of the myocardial lesion nor did they provide any information about contact of the electrode with the myocardial wall. In this study, 104 radiofrequency energy applications were performed on excised pig myocardium in circulating heparinized pig blood, and in blood alone, and root mean square (rms) voltage, current and phase angle were measured using a specially developed device. The calculated effective power and output power differed by only 2-7% measured at the point of maximum current during coagulation. A progressive drop in current following a rise in impedance led to a phase shift of more than 80 degrees with a decrease in effective power to 17% of the output power. Hence, apparent output power was mainly ineffective power. The time-dependent variations of phase angle, impedance and current were found to be useful for distinguishing between the coagulated media. These results show that physical parameters measured during radiofrequency catheter ablation may help to monitor electrode position in the clinical situation and reduce the number of ineffective energy applications.
引用
收藏
页码:62 / 68
页数:7
相关论文
共 20 条
[1]  
BELLHASSEN B, 1986, J AM COLL CARDIOL, V7, P1347
[2]   ELECTROGRAM PATTERNS PREDICTING SUCCESSFUL CATHETER ABLATION OF VENTRICULAR-TACHYCARDIA [J].
FITZGERALD, DM ;
FRIDAY, KJ ;
WAH, JAYL ;
LAZZARA, R ;
JACKMAN, WM .
CIRCULATION, 1988, 77 (04) :806-814
[3]  
FONTAINE C, 1987, CARDIAC ARRHYTHMIAS, V7, P539
[4]   CATHETER TECHNIQUE FOR CLOSED-CHEST ABLATION OF THE ATRIOVENTRICULAR-CONDUCTION SYSTEM - A THERAPEUTIC ALTERNATIVE FOR THE TREATMENT OF REFRACTORY SUPRA-VENTRICULAR TACHYCARDIA [J].
GALLAGHER, JJ ;
SVENSON, RH ;
KASELL, JH ;
GERMAN, LD ;
BARDY, GH ;
BROUGHTON, A ;
CRITELLI, G .
NEW ENGLAND JOURNAL OF MEDICINE, 1982, 306 (04) :194-200
[6]  
HINDRICKS G, 1987, Z KARDIOL S2, V70
[7]  
HOFFMANN E, 1989, 4TH EUR S CARD PAC S, V357
[8]  
HOFFMANN E, 1987, CIRCULATION S4, V278
[9]  
HOFFMANN E, 1987, Z KARDIOL S1, V58
[10]  
Hoyt RH, 1986, J APPL CARDIOL, V1, P469