LESION DETECTABILITY CONSIDERATIONS FOR AN OPTICALLY-COUPLED CCD X-RAY-IMAGING SYSTEM

被引:13
作者
LIU, H [1 ]
KARELLAS, A [1 ]
MOORE, SC [1 ]
HARRIS, LJ [1 ]
DORSI, CJ [1 ]
机构
[1] VET ADM MED CTR W ROXBURY,DEPT NUCL MED,BOSTON,MA 02132
关键词
8;
D O I
10.1109/23.322939
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
This project was aimed at investigating the lesion detectability of an optically-coupled CCD x-ray imaging system. The quasi-ideal observer signal-to-noise ratio (also called lesion-matched filter signal-to-noise ratio) was calculated for various clinical applications. This analysis differs from that of previous investigations in that we have analyzed jointly the effect of quantum noise and additive noise; we have also analyzed the combined effect of the contrast transfer property and the resolving power of the system. Numerical calculations were performed to determine lesion detectability and contrast-detail curves under mammographic and other clinical conditions. A standard mammographic phantom and a contrast-detail phantom were imaged to validate the calculation. The results demonstrate that the lesion detectability of a well designed optically-coupled CCD can be comparable to that of the state-of-the-art screen-film system under the same conditions.
引用
收藏
页码:1506 / 1509
页数:4
相关论文
共 7 条
[1]  
[Anonymous], 1983, MED IMAGING SYSTEMS
[2]  
HASEGAWWA BH, 1990, PHYSICS MED XRAY IMA
[3]   CHARGE-COUPLED DEVICE DETECTOR - PERFORMANCE CONSIDERATIONS AND POTENTIAL FOR SMALL-FIELD MAMMOGRAPHIC IMAGING APPLICATIONS [J].
KARELLAS, A ;
HARRIS, LJ ;
LIU, H ;
DAVIS, MA ;
DORSI, CJ .
MEDICAL PHYSICS, 1992, 19 (04) :1015-1023
[4]  
LIU H, 1993, SPIE P CLIN APPLICAT, V1894
[5]  
LIU H, 1992, THESIS WORECESTER PO
[7]   UNIFIED SNR ANALYSIS OF MEDICAL IMAGING-SYSTEMS [J].
WAGNER, RF ;
BROWN, DG .
PHYSICS IN MEDICINE AND BIOLOGY, 1985, 30 (06) :489-518