COMBINATION OF MICRODIALYSIS AND GLUCOSENSOR PERMITS CONTINUOUS (ON LINE) SC GLUCOSE MONITORING IN A PATIENT OPERATED DEVICE .1. INVITRO EVALUATION

被引:31
作者
KECK, FS
KERNER, W
MEYERHOFF, C
ZIER, H
PFEIFFER, EF
机构
[1] UNIV ULM,INST DIABETESTECHNOL,HELMHOLTZSTR 20,W-7900 ULM,GERMANY
[2] UNIV ULM,MED KLIN & POLIKLIN,INNERE MED ABT 1,W-7900 ULM,GERMANY
关键词
MICRODIALYSIS; ELECTROENZYMATIC GLUCOSENSOR; BIOSENSOR; GLUCOSE OXIDASE; CONTINUOUS GLUCOSE MONITORING IN INTERSTITIAL CELL FLUID;
D O I
10.1055/s-2007-1003769
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
A device for continuous glucose monitoring in fluids was obtained by combining the microdialysis technique with a measuring flow chamber of the "Glucosensor Unitec Ulm" using the GOD method for determining amperometrically blood glucose profiles. The in vitro experiments demonstrate that the relative recovery of glucose by this device is inversely related to the flow rate of the microdialysis perfusion fluid, which, in turn, is inversely related to the response time of the device. The glucose signal increases linearly with the area of the microdialysis working membrane (r = 0.98). and with the glucose concentrations of the standard solutions (r > 0.95). The variation coefficient for repeated measurements is below 8%. The accuracy of the device as demonstrated by mean measuring deviation ranges between 1 and 3.8%.
引用
收藏
页码:617 / 618
页数:2
相关论文
共 7 条
[1]   PROGRESS IN PRACTICAL ENDOCRINOLOGY - THE GLUCOSENSOR UNITEC ULM - A PORTABLE MONITOR FOR CONTINUOUS BLOOD-GLUCOSE MEASUREMENT [J].
BRUCKEL, J ;
ZIER, H ;
KERNER, W ;
PFEIFFER, EF .
HORMONE AND METABOLIC RESEARCH, 1990, 22 (07) :382-384
[2]   INVIVO MEASUREMENT OF SUBCUTANEOUS GLUCOSE-CONCENTRATIONS WITH AN ENZYMATIC GLUCOSE SENSOR AND A WICK METHOD [J].
BRUCKEL, J ;
KERNER, W ;
ZIER, H ;
STEINBACH, G ;
PFEIFFER, EF .
KLINISCHE WOCHENSCHRIFT, 1989, 67 (09) :491-495
[3]  
KECK FS, 1991, DIABETES, V40, pA476
[4]  
KERNER W, 1990, Diabetologia, V33, pA118
[5]   A MICRODIALYSIS METHOD ALLOWING CHARACTERIZATION OF INTERCELLULAR WATER SPACE IN HUMANS [J].
LONNROTH, P ;
JANSSON, PA ;
SMITH, U .
AMERICAN JOURNAL OF PHYSIOLOGY, 1987, 253 (02) :E228-E231
[6]   ARTIFICIAL BETA CELL - CONTINUOUS CONTROL OF BLOOD-SUGAR BY EXTERNAL REGULATION OF INSULIN INFUSION (GLUCOSE CONTROLLED INSULIN INFUSION SYSTEM) [J].
PFEIFFER, EF ;
THUM, C ;
CLEMENS, AH .
HORMONE AND METABOLIC RESEARCH, 1974, 6 (05) :339-342
[7]  
Zar J, 1984, BIOSTAT ANAL, V2nd