TECHNIQUE FOR MEASURING FREQUENCY-RESPONSE OF PRESSURE, VOLUME, AND FLOW TRANSDUCERS

被引:123
作者
JACKSON, AC
VINEGAR, A
机构
[1] CHILDRENS HOSP MED CTR, DEPT MED, BOSTON, MA 02115 USA
[2] HARVARD UNIV, SCH MED, DEPT PHYSIOL, BOSTON, MA 02115 USA
[3] HARVARD UNIV, SCH PUBL HLTH, DEPT PHYSIOL, BOSTON, MA 02115 USA
关键词
D O I
10.1152/jappl.1979.47.2.462
中图分类号
Q4 [生理学];
学科分类号
071003 ;
摘要
A device and methodology is presented for testing the frequency response of pressure, volume, or flow transducers. Also reported are responses of selected transducers of all three types over the range of 2-120 Hz. Several pressure transducers tested had good frequency response when connected to the test system with a minimum of interconnecting fittings; others did not. Use of additional connectors degraded the response as did the addition of air-filled catheters. The frequency responses of the pneumotachometers tested were influenced largely by the response characteristics of the associated pressure transducer and interconnecting fittings. These results emphasize the need to test the response characteristics of any transducer with specific connectors and fittings that are to be used to make the actual measurements of pressure, volume, or flow.
引用
收藏
页码:462 / 467
页数:6
相关论文
共 13 条
[1]   ACOUSTICAL IMPEDANCE OF ENCLOSURES [J].
DANIELS, FB .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1947, 19 (04) :569-571
[2]   LINEARITY AND FREQUENCY-RESPONSE OF PNEUMOTACHOGRAPHS [J].
FINUCANE, KE ;
DAWSON, SV ;
EGAN, BA .
JOURNAL OF APPLIED PHYSIOLOGY, 1972, 32 (01) :121-&
[3]   EVALUATION OF 3 TYPES OF RESPIRATORY FLOWMETERS [J].
FRY, DL ;
HYATT, RE ;
MCCALL, CB ;
MALLOS, AJ .
JOURNAL OF APPLIED PHYSIOLOGY, 1957, 10 (02) :210-214
[4]   AN EVALUATION OF MODERN PRESSURE RECORDING SYSTEMS [J].
FRY, DL ;
NOBLE, FW ;
MALLOS, AJ .
CIRCULATION RESEARCH, 1957, 5 (01) :40-46
[5]   AIRWAY GEOMETRY BY ANALYSIS OF ACOUSTIC PULSE RESPONSE MEASUREMENTS [J].
JACKSON, AC ;
BUTLER, JP ;
MILLET, EJ ;
HOPPIN, FG ;
DAWSON, SV .
JOURNAL OF APPLIED PHYSIOLOGY, 1977, 43 (03) :523-536
[6]   GLOTTAL APERTURE DURING PANTING WITH VOLUNTARY LIMITATION OF TIDAL VOLUME [J].
JACKSON, AC ;
GULESIAN, PJ ;
MEAD, J .
JOURNAL OF APPLIED PHYSIOLOGY, 1975, 39 (05) :834-836
[7]  
LANSER FJ, 1976, J APPL PHYSIOL, V41, P101
[8]   HARMONIC CONTENT OF CERTAIN RESPIRATORY FLOW PHENOMENA OF NORMAL INDIVIDUALS [J].
MCCALL, CB ;
HYATT, RE ;
NOBLE, FW ;
FRY, DL .
JOURNAL OF APPLIED PHYSIOLOGY, 1957, 10 (02) :215-218
[9]   PULMONARY MECHANICS BY SPECTRAL ANALYSIS OF FORCED RANDOM NOISE [J].
MICHAELSON, ED ;
GRASSMAN, ED ;
PETERS, WR .
JOURNAL OF CLINICAL INVESTIGATION, 1975, 56 (05) :1210-1230
[10]  
PESLIN R, 1972, B PHYSIO-PATHOL RESP, V8, P1363