HIS-BUNDLE ABLATION FOR SUPRAVENTRICULAR ARRHYTHMIAS TO AVOID SPURIOUS SHOCKS OF AN IMPLANTED DEFIBRILLATOR

被引:6
作者
JORDAENS, L
CAENEPEEL, A
CALLE, P
机构
[1] Department of Cardiology, University Hospital Ghent, Ghent
关键词
ABLATION; ATRIAL FIBRILLATION; AUTOMATIC IMPLANTABLE CARDIOVERTER DEFIBRILLATOR; VENTRICULAR TACHYCARDIA;
D O I
10.1002/clc.4960150913
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Atrial fibrillation with fast ventricular response remains a matter of concern in patients treated with an implantable cardioverter defibrillator (ICD). A patient with dilated cardiomyopathy, suffering from atrial arrhythmias and recurrent cardiac arrest due to both ventricular tachycardia and ventricular fibrillation, is presented. Ablation of the AV node by means of low-energy direct-current shocks with subsequent pacemaker implantation was performed before ICD implantation. The patient received shocks after four months, when he had recurrence of AV conduction with a slow ventricular rate. Pacemaker interaction was excluded, and no short ventricular arrhythmias were observed. During electrophysiologic study after electrical conversion of atrial fibrillation, persistent second degree heart block was documented, giving further evidence that atrial arrhythmias were not responsible for the shocks. The patient's functional status remains good after more than 18 months of follow-up.
引用
收藏
页码:693 / 695
页数:3
相关论文
共 8 条
[1]  
CLARKE M, 1991, BRIT HEART J, V66, P185
[2]   ACTUARIAL INCIDENCE AND PATTERN OF OCCURRENCE OF SHOCKS FOLLOWING IMPLANTATION OF THE AUTOMATIC IMPLANTABLE CARDIOVERTER DEFIBRILLATOR [J].
FOGOROS, RN ;
ELSON, JJ ;
BONNET, CA .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1989, 12 (09) :1465-1473
[3]  
JORDAENS L, 1988, ACTA CLIN BELG, V43, P209
[4]   LOW-DOSE ORAL SOTALOL FOR MONOMORPHIC VENTRICULAR-TACHYCARDIA - EFFECTS DURING PROGRAMMED ELECTRICAL-STIMULATION AND FOLLOW-UP [J].
JORDAENS, LJ ;
PALMER, A ;
CLEMENT, DL .
EUROPEAN HEART JOURNAL, 1989, 10 (03) :218-226
[5]  
KUCK KH, 1991, CASE STUDIE SERIES, P1
[6]   THE AUTOMATIC IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR - AN OVERVIEW [J].
MIROWSKI, M .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1985, 6 (02) :461-466
[7]   THE AUTOMATIC IMPLANTABLE CARDIOVERTER DEFIBRILLATOR - T-WAVE SENSING IN THE NEWEST GENERATION [J].
SINGER, I ;
DEBORDE, R ;
VELTRI, EP ;
SIDDOWAY, LA ;
LEVINE, JH ;
GRIFFITH, LSC ;
GUARNIERI, T .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1988, 11 (11) :1584-1591
[8]   PROGRAMMED ELECTRICAL-STIMULATION OF THE HEART IN PATIENTS WITH LIFE-THREATENING VENTRICULAR ARRHYTHMIAS - WHAT IS THE SIGNIFICANCE OF INDUCED ARRHYTHMIAS AND WHAT IS THE CORRECT STIMULATION PROTOCOL [J].
WELLENS, HJJ ;
BRUGADA, P ;
STEVENSON, WG .
CIRCULATION, 1985, 72 (01) :1-7