CALIBRATION OF THE CANON AUTOREF R-1 FOR CONTINUOUS MEASUREMENT OF ACCOMMODATION

被引:14
作者
DAVIS, B
COLLINS, M
ATCHISON, D
机构
[1] Centre for Eye Research, School of Optometry, Queensland University of Technology, Brisbane, Queensland, 4001
关键词
D O I
10.1111/j.1475-1313.1993.tb00453.x
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
The Canon Autoref R-1 is an infrared optometer which allows free-space viewing. Pugh and Winn described hardware modifications necessary to convert the Autoref R-1 from an instrument capable of taking single static measurements of the eye's accommodative (refractive) status to an instrument capable of continuous measurement of the accommodative status of the eye. We describe two methods of calibrating the Autoref R-1 for continuous measurement of accommodation for any individual eye. The assumptions and sources of error underlying these calibration techniques are discussed. We also describe further hardware modifications to simplify signal acquisition from the Autoref R-1.
引用
收藏
页码:191 / 198
页数:8
相关论文
共 8 条
[1]  
Pugh J.R., Winn B., Modification of the Canon Autoref R‐1 for use as a continuously recording infrared optometer, Ophthal. Physiol. Opt., 8, pp. 460-464, (1988)
[2]  
Pugh J.R., Winn B., A procedural guide to the modification of the Canon Autoref R‐1 for use as a continuously recording optometer, Ophthal. Physiol. Opt., 9, pp. 451-454, (1989)
[3]  
Winn B., Pugh J.R., Gilmartin B., Owens H., The effect of pupil size on static and dynamic measurements of accommodation using an infrared optometer, Ophthal. Physiol. Opt., 9, pp. 277-382, (1989)
[4]  
Matsumura I., Maruyama S., Ishikawa Y., Hirano R., Kobayashi K., Kohayakawa Y., The design of an open view autorefractometer, Advances in Diagnostic Visual Optics, pp. 36-42, (1983)
[5]  
Morgan W.W., Accommodation and its relationship to convergence, Optometry and Vision Science, 21, pp. 183-195, (1944)
[6]  
Campbell F.W., Robson J.G., Westheimer G., Fluctuations of accommodation under steady viewing, J. Physiol., 145, pp. 579-594, (1959)
[7]  
Winn B., Pugh J.R., Gilmartin B., Owens H., Arterial pulse modulates steady‐state ocular accommodation, Curr Eye Res., 9, pp. 971-975, (1990)
[8]  
Pugh J.R., Eadie A.S., Winn B., Heron G., Power spectrum analysis in the study of ocular mechanisms, Ophthal Physiol. Opt., 7, pp. 321-324, (1987)