CHARACTERIZATION OF POLYATOMIC ION INTERFERENCES IN INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETRY USING A HIGH-RESOLUTION MASS-SPECTROMETER

被引:152
作者
REED, NM [1 ]
CAIRNS, RO [1 ]
HUTTON, RC [1 ]
TAKAKU, Y [1 ]
机构
[1] MARUBUN CORP,KOHTOH KU,TOKYO 136,JAPAN
关键词
MOLECULAR INTERFERENCES; HIGH RESOLUTION INDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY;
D O I
10.1039/ja9940900881
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The effect of common interferences on the ICP-MS of the elements from Li through to Ge has been catalogued. The study utilizes the resolving power available from a double focusing mass spectrometer and provides definitive information regarding the nature and potential origin of such species. The study investigates several common anion matrices typical of the normal acids used in practical analysis (sulfate, nitrate and chloride) as well as the concomitants present from atmospheric entrainment into the argon plasma (C, N, H and O). The resulting database should provide a useful reference guide as to which interferences need to be considered in any practical analysis, whether using a quadrupole or a sector-based instrument.
引用
收藏
页码:881 / 896
页数:16
相关论文
共 20 条
[1]  
ALLENBY P, 1987, ANAL P, V24, P12
[2]  
BEAUCHEMIN D, 1989, MIKROCHIM ACTA, V111, P273
[3]   INDUCTIVELY COUPLED PLASMA AS AN ION-SOURCE FOR HIGH-RESOLUTION MASS-SPECTROMETRY [J].
BRADSHAW, N ;
HALL, EFH ;
SANDERSON, NE .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1989, 4 (08) :801-803
[4]  
CAMPARGUE R, 1966, 4 S RAR GAS DYN, P279
[5]   DETERMINATION OF RARE-EARTH ELEMENTS IN GEOLOGICAL SAMPLES BY INDUCTIVELY COUPLED PLASMA SOURCE-MASS SPECTROMETRY [J].
DATE, AR ;
HUTCHISON, D .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1987, 2 (03) :269-276
[7]  
GRAY AL, 1989, APPL INDUCTIVELY COU, P1
[8]   MASS-SPECTROMETRY OF INDUCTIVELY COUPLED PLASMAS [J].
HOUK, RS .
ANALYTICAL CHEMISTRY, 1986, 58 (01) :A97-&
[9]  
HULMSTON P, 1990, SPECTROSC INT, V3, P35
[10]  
Jarvis K. E., 1992, HDB ICP MS