SINGLE LASER 3 COLOR IMMUNOFLUORESCENCE STAINING PROCEDURES BASED ON ENERGY-TRANSFER BETWEEN PHYCOERYTHRIN AND CYANINE-5

被引:42
作者
LANSDORP, PM
SMITH, C
SAFFORD, M
TERSTAPPEN, LWMM
THOMAS, TE
机构
[1] UNIV BRITISH COLUMBIA,DEPT MED,VANCOUVER V6T 1W5,BC,CANADA
[2] BECTON DICKINSON IMMUNOCYTOMETRY SYST,SAN JOSE,CA
来源
CYTOMETRY | 1991年 / 12卷 / 08期
关键词
FLUORESCENT TANDEM CONJUGATES; TETRAMERIC ANTIBODY COMPLEXES; FLOW CYTOMETRY; ANTIBODY LABELING; ANTIBODY-ANTIGEN INTERACTION;
D O I
10.1002/cyto.990120806
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Monoclonal antibodies specific for phycoerythrin (PE) were covalently labeled with the fluorescent dye cyanine 5 (Cy5). Excitation at 488 nm of immune complexes obtained by mixing Cy5-anti-PE with PE resulted in a 4-fold reduction of PE fluorescence measured at 565 nm and an increase of fluorescence measured at 655 nm. The observed energy transfer between PE and Cy5-anti-PE was used to develop three color immunofluorescence staining procedures for flow cytometers equipped with an Argon laser tuned at 488nm. Mouse IgG1 monoclonal antibodies specific for cell surface antigens were cross-linked with either unlabeled or Cy5 labeled mouse IgG1 anti-PE using F(ab'), fragments of monoclonal rat anti-mouse IgG1. PE was added to these immune complexes in sufficient amounts to saturate all PE binding sites. Cells were incubated with PE-labeled and PE/Cy5-labeled tetrameric antibody complexes together with FITC labeled antibodies and analyzed by flow cytometry. The emission from FITC, PE and PE/ Cy5 could be readily separated and bright three color immunofluorescence staining of mononuclear cells from human peripheral blood and bone marrow was observed. The results of these experiments demonstrate that useful probes for single laser three color staining of cell surface antigens can be readily obtained by mixing of selected reagents. Compared to standard procedures for the covalent labeling of PE (tandem) molecules to antibodies, the non-covalent procedures described in this report provide significant advantages in terms of the amount of reagents, time and equipment required to obtain suitable reagents for three color immunofluorescence staining.
引用
收藏
页码:723 / 730
页数:8
相关论文
共 9 条
[1]   FLUORESCENT TANDEM PHYCOBILIPROTEIN CONJUGATES - EMISSION WAVELENGTH SHIFTING BY ENERGY-TRANSFER [J].
GLAZER, AN ;
STRYER, L .
BIOPHYSICAL JOURNAL, 1983, 43 (03) :383-386
[2]  
Hardy RR, 1986, HDB EXPT IMMUNOLOGY, V1
[3]   AN IMPROVED FLOW CYTOMETRIC ASSAY FOR SOMATIC MUTATIONS AT THE GLYCOPHORIN-A LOCUS IN HUMANS [J].
LANGLOIS, RG ;
NISBET, BA ;
BIGBEE, WL ;
RIDINGER, DN ;
JENSEN, RH .
CYTOMETRY, 1990, 11 (04) :513-521
[4]   PURIFICATION AND ANALYSIS OF BISPECIFIC TETRAMERIC ANTIBODY COMPLEXES [J].
LANSDORP, PM ;
THOMAS, TE .
MOLECULAR IMMUNOLOGY, 1990, 27 (07) :659-666
[5]   SELECTIVE EXPRESSION OF CD45 ISOFORMS ON FUNCTIONAL SUBPOPULATIONS OF CD34+ HEMATOPOIETIC-CELLS FROM HUMAN BONE-MARROW [J].
LANSDORP, PM ;
SUTHERLAND, HJ ;
EAVES, CJ .
JOURNAL OF EXPERIMENTAL MEDICINE, 1990, 172 (01) :363-366
[6]   FLUORESCENT PHYCOBILIPROTEIN CONJUGATES FOR ANALYSES OF CELLS AND MOLECULES [J].
OI, VT ;
GLAZER, AN ;
STRYER, L .
JOURNAL OF CELL BIOLOGY, 1982, 93 (03) :981-986
[7]   CYANINE DYE LABELING REAGENTS - CARBOXYMETHYLINDOCYANINE SUCCINIMIDYL ESTERS [J].
SOUTHWICK, PL ;
ERNST, LA ;
TAURIELLO, EW ;
PARKER, SR ;
MUJUMDAR, RB ;
MUJUMDAR, SR ;
CLEVER, HA ;
WAGGONER, AS .
CYTOMETRY, 1990, 11 (03) :418-430
[8]  
STRYLER L, 1985, Patent No. 4542104
[9]   USE OF TETRAMERIC ANTIBODY COMPLEXES TO STAIN CELLS FOR FLOW-CYTOMETRY [J].
WOGNUM, AW ;
THOMAS, TE ;
LANSDORP, PM .
CYTOMETRY, 1987, 8 (04) :366-371