USE OF SENTINEL BLOOD CULTURE SYSTEM FOR ANALYSIS OF SPECIMENS FROM POTENTIALLY INFECTED PROSTHETIC JOINTS

被引:2
作者
BAKER, S [1 ]
FRAISE, AP [1 ]
机构
[1] SELLY OAK HOSP,DEPT MICROBIOL,BIRMINGHAM B29 6JD,ENGLAND
关键词
D O I
10.1136/jcp.47.5.475
中图分类号
R36 [病理学];
学科分类号
100104 ;
摘要
The Sentinel blood culture system was used for the analysis of 657 specimens from infected prosthetic joints and blood cultures (83 from prosthetic joints and 574 from standard blood cultures). The positivity rate was similar for specimens from prosthetic joints and blood cultures (18% compared with 14%). However, there was an unacceptable rate of false positive results with specimens from prosthetic joints (58% compared with 8%). This high false positivity rate was due to (i) prolonged incubation and (ii) the lack of blood in these specimens. It is therefore recommended that the Sentinel system should only be used for the initial seven days of incubation of specimens taken from prosthetic joints. Further incubation should take place in a standard incubator and a terminal subculture performed after 21 days.
引用
收藏
页码:475 / 476
页数:2
相关论文
共 6 条
  • [1] ACAR JF, 1991, ANTIBIOTICS LABORATO, P47
  • [2] CLINICAL AND MICROBIAL FEATURES OF PROSTHETIC JOINT INFECTION
    INMAN, RD
    GALLEGOS, KV
    BRAUSE, BD
    REDECHA, PB
    CHRISTIAN, CL
    [J]. AMERICAN JOURNAL OF MEDICINE, 1984, 77 (01) : 47 - 53
  • [3] DIAGNOSING INFECTION IN HIP REPLACEMENTS - THE USE OF FINE-NEEDLE ASPIRATION AND RADIOMETRIC CULTURE
    ROBERTS, P
    WALTERS, AJ
    MCMINN, DJW
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1992, 74 (02): : 265 - 269
  • [4] SLACK MPE, 1990, TOPLEY WILSONS PRINC, V2, P358
  • [5] Whyte W, 1981, J Hosp Infect, V2, P219, DOI 10.1016/0195-6701(81)90041-4
  • [6] 1991, MDD9139 MED DEV DIR, P10