CATHETER-IMPLANTED PROSTHETIC HEART-VALVES - TRANSLUMINAL CATHETER IMPLANTATION OF A NEW EXPANDABLE ARTIFICIAL-HEART VALVE IN THE DESCENDING THORACIC AORTA IN ISOLATED VESSELS AND CLOSED-CHEST PIGS

被引:13
作者
KNUDSEN, LL
ANDERSEN, HR
HASENKAM, JM
机构
[1] AARHUS UNIV HOSP, SKEJBY SYGEHUS, DEPT CARDIOL, DK-8200 AARHUS N, DENMARK
[2] AARHUS UNIV HOSP, SKEJBY SYGEHUS, DEPT THORAC & CARDIOVASC SURG, DK-8200 AARHUS N, DENMARK
[3] AARHUS UNIV, INST EXPTL CLIN RES, DK-8000 AARHUS, DENMARK
[4] AARHUS UNIV, CARDIOVASC RES CTR, DK-8000 AARHUS, DENMARK
关键词
HEART VALVE PROSTHESIS; STENT-VALVE; TRANSLUMINAL IMPLANTATION; PIGS;
D O I
10.1177/039139889301600505
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A new expandable artificial heart valve was developed for implantation by a transluminal catheter technique without using thoracotomy or extracorporeal circulation. The aim of this study was to implant the valve in isolated vessels of the descending thoracic aorta as well as in closed chest pigs, and furthermore to study the prosthesis' mechanical stability and the valve function. The artificial valve was made by mounting a porcine aortic valve on an expandable stent. Before implantation, the stent-valve was compressed on a deflated balloon catheter and mounted inside an introducer sheath. After intravascular introduction to the descending thoracic aorta the stent-valve was discharged from the sheath. Implantation was performed by balloon inflation which expanded the stent-valve to a diameter exceeding the internal vessel's diameter. After balloon deflation the stent-valve maintained an expanded configuration ensuring a stable fixation against the vessel wall. In vitro implantations were performed in 36 isolated descending thoracic aorta specimens obtained from 80 kg pigs. Mechanical stability was evaluated by applying a downing load to the prosthesis. No displacement occurred at loads less-than-or-equal-to 1 kg when a large balloon (31 mm) was used for implantation. Transvalvular pressure differences between 11-47 mmHg (median) were obtained at antegrade flowrates between 5-8 l/min. Furthermore, only moderate leakage flows were measured during retrograde perfusion. In vivo implantations were performed in six 80 kg pigs. Implantation was safe and easy, and angiograph and haemodynamic evaluations revealed essentially no stenosis or regurgitation. No complications in migration, perforation, hemorrhage or thrombosis were observed. This study indicates a good mechanical stability and valve function of the new expandable artificial valves.
引用
收藏
页码:253 / 262
页数:10
相关论文
共 18 条
[1]   TRANSLUMINAL IMPLANTATION OF ARTIFICIAL-HEART VALVES - DESCRIPTION OF A NEW EXPANDABLE AORTIC-VALVE AND INITIAL RESULTS WITH IMPLANTATION BY CATHETER TECHNIQUE IN CLOSED CHEST PIGS [J].
ANDERSEN, HR ;
KNUDSEN, LL ;
HASENKAM, JM .
EUROPEAN HEART JOURNAL, 1992, 13 (05) :704-708
[2]  
CONKLIN WS, 1958, J THORAC SURG, V36, P238
[3]   PARTIAL AND COMPLETE PROSTHESES IN AORTIC INSUFFICIENCY [J].
HARKEN, DE ;
SOROFF, HS ;
TAYLOR, WJ ;
LEFEMINE, AA ;
GUPTA, SK ;
LUNZER, S .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1960, 40 (06) :744-762
[4]   A MODEL FOR ACUTE HEMODYNAMIC-STUDIES IN THE ASCENDING AORTA IN PIGS [J].
HASENKAM, JM ;
OSTERGAARD, JH ;
PEDERSEN, EM ;
PAULSEN, PK ;
NYGAARD, H ;
SCHURIZEK, BA ;
JOHANNSEN, G .
CARDIOVASCULAR RESEARCH, 1988, 22 (07) :464-471
[5]  
HILBERT SL, 1987, J THORAC CARDIOV SUR, V94, P419
[6]  
HUFNAGEL CA, 1979, AM J SURG, V137, P285, DOI 10.1016/0002-9610(79)90054-0
[7]  
HUFNAGEL CA, 1954, SURGERY, V35, P673
[8]  
HUFNAGEL CA, 1953, B GEORGETOWN U MED C, V6, P3
[9]  
LO HB, 1988, T AM SOC ART INT ORG, V34, P839
[10]  
REUL H, 1986, P HEART VALVE ENG, P3