GEL-PERMEATION CHROMATOGRAPHY FOURIER-TRANSFORM INFRARED INTERFACE FOR POLYMER ANALYSIS

被引:38
作者
WHEELER, LM [1 ]
WILLIS, JN [1 ]
机构
[1] LAB CONNECT INC,MARLBOROUGH,MA 01752
关键词
ANALYTICAL METHODS; GP FT-IR;
D O I
10.1366/0003702934067982
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A new solvent elimination interface capable of operating at elevated temperatures, here 145-degrees-C, has been used to collect polymer molecular weight fractions eluting from a gel permeation chromatogram and to prepare them for IR analysis. The sample is deposited continuously onto a rotating germanium disk which can subsequently be scanned with the use of GC/FT-IR software, allowing direct access to the polymer or copolymer composition as a function of molecular weight. Data are presented here for an ethylene-propylene copolymer which has a distinct bimodal molecular weight distribution. Both the concentration profile and the ''composition distribution'' are examined. For the polymer concentration profile, comparison is made between the chromatogram obtained with a differential refractive index (DRI) detector and the IR detector (plotting the absorbance as a function of time using Gram-Schmidt vector orthogonalization). The copolymer composition is determined from the relative absorbance of methyl and methylene groups in the CH stretching region. The results show a small change in propylene content as a function of molecular weight, and there is good agreement between composition calculated with the use of the Gram-Schmidt and point-to-point methods.
引用
收藏
页码:1128 / 1130
页数:3
相关论文
共 6 条
[1]  
Datta S., 1988, U.S. Patent, Patent No. [4,722,971, 4722971]
[2]   INTERFACE TO ELIMINATE HIGH-BOILING GEL-PERMEATION CHROMATOGRAPHIC SOLVENTS ONLINE FOR POLYMER COMPOSITION DRIFT STUDIES [J].
DEKMEZIAN, AH ;
MORIOKA, T .
ANALYTICAL CHEMISTRY, 1989, 61 (05) :458-461
[3]   CONTINUOUS INFRARED SPECTROSCOPIC ANALYSIS OF ISOCRATIC AND GRADIENT ELUTION REVERSED-PHASE LIQUID-CHROMATOGRAPHY SEPARATIONS [J].
GAGEL, JJ ;
BIEMANN, K .
ANALYTICAL CHEMISTRY, 1987, 59 (09) :1266-1272
[4]   NOVEL APPROACHES TO INTERFACING A HIGH-PERFORMANCE LIQUID CHROMATOGRAPH WITH A FOURIER-TRANSFORM INFRARED SPECTROMETER [J].
KUEHL, D ;
GRIFFITHS, PR .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1979, 17 (08) :471-476
[5]   MAGIC-LC/FT-IR SPECTROMETRY - PRELIMINARY STUDIES [J].
ROBERTSON, RM ;
DEHASETH, JA ;
KIRK, JD ;
BROWNER, RF .
APPLIED SPECTROSCOPY, 1988, 42 (08) :1365-1368
[6]  
VIDRINE W, 1975, MOL SPECTROSC, V32, P502