THE POINT-SPREAD FUNCTION OF A CONFOCAL MICROSCOPE - ITS MEASUREMENT AND USE IN DECONVOLUTION OF 3-D DATA

被引:126
作者
SHAW, PJ
RAWLINS, DJ
机构
[1] Department of Cell Biology, John Innes Institute, Norwich, NR4 7UH, Colney Lane
关键词
CONFOCAL MICROSCOPY; 3-DIMENSIONAL FLUORESCENCE MICROSCOPY; POINT-SPREAD FUNCTION; DECONVOLUTION; COMPUTER IMAGE PROCESSING;
D O I
10.1111/j.1365-2818.1991.tb03168.x
中图分类号
TH742 [显微镜];
学科分类号
摘要
We have measured the point-spread function (PSF) for an MRC-500 confocal scanning laser microscope using subresolution fluorescent beads. PSFs were measured for two lenses of high numerical aperture - the Zeiss plan-neofluar 63 x water immersion and Leitz plan-apo 63 x oil immersion - at three different sizes of the confocal detector aperture. The measured PSFs are fairly symmetrical, both radially and axially. In particular there is considerably less axial asymmetry than has been demonstrated in measurements of conventional (non-confocal) PSFs. Measurements of the peak width at half-maximum peak height for the minimum detector aperture gave approximately 0.23 and 0.8-mu-m for the radial and axial resolution respectively (4.6 and 15.9 in dimensionless optical units). This increased to 0.38 and 1.5-mu-m (7.5 and 29.8 in dimensionless units) for the largest detector aperture examined. The resulting optical transfer functions (OTFs) were used in an iterative, constrained deconvolution procedure to process three-dimensional confocal data sets from a biological specimen - pea root cells labelled in situ with a fluorescent probe to ribosomal genes. The deconvolution significantly improved the clarity and contrast of the data. Furthermore, the loss in resolution produced by increasing the size of the detector aperture could be restored by the deconvolution procedure. Therefore for many biological specimens which are only weakly fluorescent it may be preferable to open the detector aperture to increase the strength of the detected signal, and thus the signal-to-noise ratio, and then to restore the resolution by deconvolution.
引用
收藏
页码:151 / 165
页数:15
相关论文
共 20 条
[1]  
AGARD DA, 1989, METHOD CELL BIOL, V30, P353
[2]  
AGARD DA, 1984, ANNU REV BIOPHYS BIO, V13, P191
[3]   3-DIMENSIONAL ARCHITECTURE OF A POLYTENE NUCLEUS [J].
AGARD, DA ;
SEDAT, JW .
NATURE, 1983, 302 (5910) :676-681
[4]  
Born M., 1964, PRINCIPLES OPTICS
[5]  
Castleman K. R., 1979, DIGITAL IMAGE PROCES
[6]  
CONAN V, 1990, T ROY MICR, V1, P281
[7]   RECONSTRUCTING 3-D LIGHT-MICROSCOPIC IMAGES BY DIGITAL IMAGE-PROCESSING [J].
ERHARDT, A ;
ZINSER, G ;
KOMITOWSKI, D ;
BILLE, J .
APPLIED OPTICS, 1985, 24 (02) :194-200
[8]   DETERMINATION OF 3-DIMENSIONAL IMAGING PROPERTIES OF A LIGHT-MICROSCOPE SYSTEM - PARTIAL CONFOCAL BEHAVIOR IN EPIFLUORESCENCE MICROSCOPY [J].
HIRAOKA, Y ;
SEDAT, JW ;
AGARD, DA .
BIOPHYSICAL JOURNAL, 1990, 57 (02) :325-333
[9]   MAXIMUM-LIKELIHOOD IMAGE-RESTORATION ADAPTED FOR NONCOHERENT OPTICAL IMAGING [J].
HOLMES, TJ .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1988, 5 (05) :666-673
[10]  
JANNSON PA, 1984, DECONVOLUTION APPLIC