ENERGY CALIBRATION OF A 5.5 MV VANDEGRAAFF ACCELERATOR USING A TIME-OF-FLIGHT TECHNIQUE

被引:14
作者
ANDRADE, E
FEREGRINO, M
ZAVALA, EP
PINEDA, JC
JIMENEZ, R
JAIDAR, A
机构
[1] Instituto de Fisica, Universidad Nacional Autonoma de Mexico, México 20, DF CP 01000
关键词
D O I
10.1016/0168-9002(90)91784-9
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Energy calibration of the analysing magnet for the University of México 5.5 MV Van de Graaff accelerator has been carried out using a time-of-flight method. The pulsed and bunched beam produced by the accelerator was used to deduce the energy E of the beam particles from the transit time between a coaxial pickup and a charged particle detector separated by 1.5 m. The measured FWHM energy resolution of this system was about 20 keV. © 1990.
引用
收藏
页码:135 / 138
页数:4
相关论文
共 4 条