EFFECT OF MOBILE PHASE DENSITY GRADIENTS ON EFFICIENCY IN PACKED-COLUMN SUPERCRITICAL FLUID CHROMATOGRAPHY

被引:29
作者
BERGER, TA
DEYE, JF
机构
[1] Hewlett-Packard Company, Avondale, 19311-0900, PA
关键词
Density gradients; Effect on efficiency; Packed columns; Supercritical fluid chromatography; (SFC);
D O I
10.1007/BF02270449
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The efficiency of packed columns was measured as a function of flow rate, temperature, outlet density, and the density differential across the column, unsing pure carbon dioxide as the mobile phase. Although density differentials are often blamed for a serious loss in efficiency in packed column supercritical fluid chromatography, the results show that efficiency was not a function of the density differential. Peak shapes suggest that apparent loss in efficiency is actually due to inadequate solubility of the solute in carbon dioxide. © 1990 Friedr. Vieweg & Sohn Verlagsgesellschaft mbH.
引用
收藏
页码:57 / 60
页数:4
相关论文
共 15 条
  • [1] BENDER E, 1970, 5TH P S THERM PROP N
  • [2] SUPERCRITICAL FLUID CHROMATOGRAPHY WITH SMALL PARTICLE DIAMETER PACKED-COLUMNS
    GERE, DR
    BOARD, R
    MCMANIGILL, D
    [J]. ANALYTICAL CHEMISTRY, 1982, 54 (04) : 736 - 740
  • [3] EFFECTS OF COLUMN LENGTH, PARTICLE-SIZE, FLOW-RATE, AND PRESSURE PROGRAMMING RATE ON RESOLUTION IN PRESSURE-PROGRAMMED SUPERCRITICAL FLUID CHROMATOGRAPHY
    GRAHAM, JA
    ROGERS, LB
    [J]. JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1980, 18 (02) : 75 - 84
  • [4] ISOCRATIC NETWORKS IN SUPERCRITICAL FLUID CHROMATOGRAPHY .2. SELECTIVITIES AND EFFECTIVE PLATE NUMBERS
    LEYENDECKER, D
    LEYENDECKER, D
    SCHMITZ, FP
    KLESPER, E
    [J]. JOURNAL OF HIGH RESOLUTION CHROMATOGRAPHY & CHROMATOGRAPHY COMMUNICATIONS, 1987, 10 (03): : 141 - 144
  • [5] SUPERCRITICAL FLUID CHROMATOGRAPHY USING MIXTURES OF CARBON-DIOXIDE OR ETHANE WITH 1,4-DIOXANE AS ELUENTS
    LEYENDECKER, D
    LEYENDECKER, D
    SCHMITZ, FP
    LORENSCHAT, B
    KLESPER, E
    [J]. JOURNAL OF CHROMATOGRAPHY, 1987, 398 : 105 - 123
  • [6] 3-DIMENSIONAL NETWORK PLOTS IN SUPERCRITICAL FLUID CHROMATOGRAPHY WITH NORMAL-PENTANE-1,4-DIOXANE
    LEYENDECKER, D
    SCHMITZ, FP
    LEYENDECKER, D
    KLESPER, E
    [J]. JOURNAL OF CHROMATOGRAPHY, 1987, 393 (02): : 155 - 167
  • [7] CHROMATOGRAPHY WITH SUBCRITICAL AND SUPERCRITICAL ELUENTS - INFLUENCE OF TEMPERATURE, PRESSURE AND FLOW-RATE ON THE BEHAVIOR OF LOWER ALKANES
    LEYENDECKER, D
    SCHMITZ, FP
    LEYENDECKER, D
    KLESPER, E
    [J]. JOURNAL OF CHROMATOGRAPHY, 1985, 321 (02): : 273 - 286
  • [8] INFLUENCES OF TEMPERATURE, PRESSURE AND ELUENT COMPOSITION ON THE BEHAVIOR OF BINARY ELUENTS AT SUBCRITICAL AND SUPERCRITICAL CONDITIONS - PENTANE AND DIETHYLENEGLYCOL DIMETHYL ETHER
    LEYENDECKER, D
    SCHMITZ, FP
    KLESPER, E
    [J]. CHROMATOGRAPHIA, 1987, 23 (03) : 171 - 178
  • [9] CHROMATOGRAPHY WITH SUBCRITICAL AND SUPERCRITICAL ELUENTS - THE INFLUENCE OF TEMPERATURE, PRESSURE AND FLOW-RATE ON THE BEHAVIOR OF DIMETHYL AND DIETHYL-ETHER
    LEYENDECKER, D
    SCHMITZ, FP
    KLESPER, E
    [J]. JOURNAL OF CHROMATOGRAPHY, 1984, 315 (DEC): : 19 - 30
  • [10] THE DEPENDENCE OF REDUCED PLATE HEIGHT ON REDUCED VELOCITY IN CARBON-DIOXIDE SUPERCRITICAL FLUID CHROMATOGRAPHY WITH PACKED-COLUMNS
    MOURIER, PA
    CAUDE, MH
    ROSSET, RH
    [J]. CHROMATOGRAPHIA, 1987, 23 (01) : 21 - 25