INTRACRANIAL-PRESSURE MONITORING IN CHILDREN - COMPARISON OF EXTERNAL VENTRICULAR DEVICE WITH THE FIBEROPTIC SYSTEM

被引:32
作者
GAMBARDELLA, G
ZACCONE, C
CARDIA, E
TOMASELLO, F
机构
[1] POLICLIN UNIV, NEUROCHIRURG CLIN, I-98100 MESSINA, ITALY
[2] POLICLIN UNIV, SEZ NEUROCHIRURG INFANTILE, I-98100 MESSINA, ITALY
关键词
INTRACRANIAL PRESSURE MONITORING; FIBEROPTIC DEVICE; VENTRICULAR DEVICE; CRANIOCEREBRAL TRAUMA; CHILDREN;
D O I
10.1007/BF00393552
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Several intracranial pressure monitoring devices have been developed in the past several years. We have recently adopted the Camino fiberoptic device that permits subdural, intraparenchymal, and intraventricular monitoring. In this report we compare experiences in monitoring a group of pediatric patients with severe craniocerebral trauma and coma, grouped according to severity of Glasgow Coma Scale score. Patient age ranged from 2 to 16 years. Twelve patients were monitored by a ventricular catheter and 37, treated more recently, by a Camino fiberoptic device. The study demonstrated that the fiberoptic device and the ventricular catheter have the same accuracy and reliability. The fiberoptic method correlates very closely with the ventriculostomy method, but the pressure values are always 3 +/- 2 mmHg lower than those obtained with the conventional pressure transducer system, especially in more critically ill patients. This new technique is also easier to implant, safer to use, has minimal drift, and is minimally invasive, which particularly speaks for its use in pediatric patients.
引用
收藏
页码:470 / 473
页数:4
相关论文
共 16 条
[2]   INTRACRANIAL-PRESSURE MONITORS - EPIDEMIOLOGIC-STUDY OF RISK-FACTORS AND INFECTIONS [J].
AUCOIN, PJ ;
KOTILAINEN, HR ;
GANTZ, NM ;
DAVIDSON, R ;
KELLOGG, P ;
STONE, B .
AMERICAN JOURNAL OF MEDICINE, 1986, 80 (03) :369-376
[3]   CLINICAL-EVALUATION OF 2 METHODS OF SUBDURAL PRESSURE MONITORING [J].
BARLOW, P ;
MENDELOW, AD ;
LAWRENCE, AE ;
BARLOW, M ;
ROWAN, JO .
JOURNAL OF NEUROSURGERY, 1985, 63 (04) :578-582
[4]  
Choux M, 1986, HEAD INJURIES NEWBOR, P163
[5]   MEASUREMENT OF EXTRADURAL PRESSURE AND ITS RELATIONSHIP TO OTHER INTRACRANIAL PRESSURES - EXPERIMENTAL AND CLINICAL STUDY [J].
CORONEOS, NJ ;
MCDOWALL, DG ;
GIBSON, RM ;
PICKERODT, VW ;
KEANEY, NP .
JOURNAL OF NEUROLOGY NEUROSURGERY AND PSYCHIATRY, 1973, 36 (04) :514-522
[6]   EVALUATION OF A FIBEROPTIC INTRACRANIAL-PRESSURE MONITOR [J].
CRUTCHFIELD, JS ;
NARAYAN, RK ;
ROBERTSON, CS ;
MICHAEL, LH .
JOURNAL OF NEUROSURGERY, 1990, 72 (03) :482-487
[7]   ASSESSMENT OF LEEDS DEVICE FOR MONITORING INTRACRANIAL-PRESSURE [J].
DEARDEN, NM ;
MCDOWALL, DG ;
GIBSON, RM .
JOURNAL OF NEUROSURGERY, 1984, 60 (01) :123-129
[8]  
GAMBARDELLA G, 1991, 8 INT S INTR PRESS R
[9]   INTRACRANIAL RUPTURE OF A PRESSURE MONITORING TRANSDUCER - TECHNICAL NOTE [J].
GENTLEMAN, D ;
MENDELOW, AD .
NEUROSURGERY, 1986, 19 (01) :91-92
[10]  
GUILLAUME J, 1951, PRESSE MED, V59, P953