ABSORBABLE INTRAMEDULLARY IMPLANTS FOR HAND FRACTURES - ANIMAL-EXPERIMENTS AND CLINICAL-TRIAL

被引:40
作者
KUMTA, SM
SPINNER, R
LEUNG, PC
机构
来源
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME | 1992年 / 74卷 / 04期
关键词
D O I
10.1302/0301-620X.74B4.1320621
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Biodegradable implants made from polyglycolic and polylactic acid co-polymers undergo degradation by hydrolysis which results in loss of their mechanical strength. The degradation of 1.5 mm polyglycolide rods (Biofix) was studied after intramedullary and subcutaneous implantation in rabbits. Two weeks after implantation there was a 73% reduction in strength of the intramedullary implants and a 64% reduction in the subcutaneous implants. Polyglycolide implants were compared with Kirschner wires for intramedullary fixation of extra-articular fractures in the hand. In one group of patients fractures were fixed with a 1.5 mm intramedullary rod and in a similar group a Kirschner wire was used. In both a wire loop was added for extra fixation. At six months there was no significant difference between the two groups. There were no allergic reactions to the polyglycolide implants.
引用
收藏
页码:563 / 566
页数:4
相关论文
共 6 条
[1]   PRELIMINARY-REPORT ON THE OSTEOGENIC POTENTIAL OF A BIODEGRADABLE CO-POLYMER OF POLYACTIDE (PLA) AND POLYGLYCOLIDE (PGA) [J].
HOLLINGER, JO .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1983, 17 (01) :71-82
[2]   COMBINED INTRAMEDULLARY KIRSCHNER WIRE AND INTRA-OSSEOUS WIRE LOOP FOR FIXATION OF FINGER FRACTURES [J].
HUNG, LK ;
SO, WS ;
LEUNG, PC .
JOURNAL OF HAND SURGERY-BRITISH AND EUROPEAN VOLUME, 1989, 14B (02) :171-176
[3]  
MAKELA EA, 1987, J PEDIATR ORTHOPED, V7, P415
[4]  
VAINIONPAA S, 1986, ARCH ORTHOP TRAUM SU, V104, P333, DOI 10.1007/BF00454425
[5]   STRENGTH AND STRENGTH RETENTION INVITRO, OF ABSORBABLE, SELF-REINFORCED POLYGLYCOLIDE (PGA) RODS FOR FRACTURE FIXATION [J].
VAINIONPAA, S ;
KILPIKARI, J ;
LAIHO, J ;
HELEVIRTA, P ;
ROKKANEN, P ;
TORMALA, P .
BIOMATERIALS, 1987, 8 (01) :46-48
[6]  
WILLIAMS DF, 1979, SPECH TECH PUBL, V684, P61