DETERMINATION OF TRACE AMOUNTS OF SILVER WITH A CHEMICALLY MODIFIED CARBON PASTE ELECTRODE

被引:32
作者
LI, PB [1 ]
GAO, ZQ [1 ]
XU, YW [1 ]
WANG, GQ [1 ]
ZHAO, ZF [1 ]
机构
[1] WUHAN UNIV,DEPT CHEM,WUHAN,PEOPLES R CHINA
关键词
D O I
10.1016/S0003-2670(00)85131-4
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A method for the determination of trace amounts of silver with a chemically modified carbon paste electrode is described. The modified electrode is prepared by simply mixing a chelating resin (a polythioether backbone and dioxymonosulphur polyethylene polyimines in the side-chain polymer) with graphite powder and Nujol oil. By immersing the electrode in a silver sample solution (pH = 6.5-7.5), silver can be adsorbed on the electrode surface and then determined by voltammetry in a separate blank solution. The response depends on the concentration of silver and the preconcentration time. For a preconcentration time of 5 min, the detection limit is about 3 × 10-10 M and the linear range is from 5 × 10-10 to 1 × 10-7 M with a relative standard deviation of 4%. Many common metal ions have no or little effect on the determination of silver. The recommended procedure was applied to the determination of trace amounts of silver in waste water. © 1990.
引用
收藏
页码:213 / 219
页数:7
相关论文
共 23 条